Javascript must be enabled to continue!
Corneal complication after femtosecond laser-assisted cataract surgery
View through CrossRef
Abstract
Rationale:
The femtosecond laser LDV Z 8 is unique, and the only femtosecond laser used in ophthalmic microsurgery, which is characterized by low-energy near-infrared (1030 nm) femtosecond single pulses in the nano-Joule range and a very high repetition rate in the MHz range. To the best of our knowledge, this is the first description of unintentional partial corneal incision in the anterior part of a stroma by the femtosecond laser.
Patient concerns:
A 79-year-old patient was referred to our clinic for cataract surgery. On admission, we diagnosed mature cataract of the left eye and pseudophakic of the right eye. The patient was qualified for the Femtosecond Laser-Assisted Cataract Surgery (FLACS). Unexpectedly during FLACS procedure after lens fragmentation the surgeon observed unexpected paracentral incision in the cornea.
Diagnosis:
The corneal incision line between 4 to 8 o’clock was observed. In vivo confocal microscopy was performed to assess the morphology and depth of the corneal wound. The corneal incision covered the epithelium, Bowman's membrane and stroma of the cornea reached a depth of 336 um.
Interventions:
The patient was under increased ophthalmologic controls; follow up with typical ophthalmic examinations and confocal microscopy was performed.
Outcomes:
In 2-year follow-up period, this complication had no effect on postoperative visual function, the patient had no visual problems and obtained the final BCVA 5/5. There is no dislocation of the lens in 2 years follow-up.
Lessons:
Low pulse energy and high pulse frequency in the LDV Z8 causes a low traumatization of tissues. In a 2-year follow-up, the corneal incision line could be observed on the slit lamp examination without long-term visual consequences of this complication. In our opinion, the most likely cause of this complication was human error and lack of communication between medical personnel. The operation team should be alert and great caution must be exercised during the procedure and check the laser settings parameters carefully each time.
Ovid Technologies (Wolters Kluwer Health)
Title: Corneal complication after femtosecond laser-assisted cataract surgery
Description:
Abstract
Rationale:
The femtosecond laser LDV Z 8 is unique, and the only femtosecond laser used in ophthalmic microsurgery, which is characterized by low-energy near-infrared (1030 nm) femtosecond single pulses in the nano-Joule range and a very high repetition rate in the MHz range.
To the best of our knowledge, this is the first description of unintentional partial corneal incision in the anterior part of a stroma by the femtosecond laser.
Patient concerns:
A 79-year-old patient was referred to our clinic for cataract surgery.
On admission, we diagnosed mature cataract of the left eye and pseudophakic of the right eye.
The patient was qualified for the Femtosecond Laser-Assisted Cataract Surgery (FLACS).
Unexpectedly during FLACS procedure after lens fragmentation the surgeon observed unexpected paracentral incision in the cornea.
Diagnosis:
The corneal incision line between 4 to 8 o’clock was observed.
In vivo confocal microscopy was performed to assess the morphology and depth of the corneal wound.
The corneal incision covered the epithelium, Bowman's membrane and stroma of the cornea reached a depth of 336 um.
Interventions:
The patient was under increased ophthalmologic controls; follow up with typical ophthalmic examinations and confocal microscopy was performed.
Outcomes:
In 2-year follow-up period, this complication had no effect on postoperative visual function, the patient had no visual problems and obtained the final BCVA 5/5.
There is no dislocation of the lens in 2 years follow-up.
Lessons:
Low pulse energy and high pulse frequency in the LDV Z8 causes a low traumatization of tissues.
In a 2-year follow-up, the corneal incision line could be observed on the slit lamp examination without long-term visual consequences of this complication.
In our opinion, the most likely cause of this complication was human error and lack of communication between medical personnel.
The operation team should be alert and great caution must be exercised during the procedure and check the laser settings parameters carefully each time.
Related Results
<b>Corneal Biomechanical Integrity After Femtosecond Laser-Assisted Versus Manual Incisions in Cataract Surgery: A Systematic Review</b>
<b>Corneal Biomechanical Integrity After Femtosecond Laser-Assisted Versus Manual Incisions in Cataract Surgery: A Systematic Review</b>
Background: The integrity of the clear corneal incision is fundamental to safety, wound sealing, and refractive stability in cataract surgery. Although femtosecond laser-assisted c...
Changes in Aqueous Humor pH After Femtosecond Laser-Assisted Cataract Surgery
Changes in Aqueous Humor pH After Femtosecond Laser-Assisted Cataract Surgery
PURPOSE:
To compare aqueous humor pH values in patients during femtosecond laser-assisted cataract surgery with patients during conventional phacoemulsification.
...
“Prevalence And Determinants Of Age-Related Cataract In The Rural Population: A Cross-Sectional Study”
“Prevalence And Determinants Of Age-Related Cataract In The Rural Population: A Cross-Sectional Study”
Background of the Study: Age-related cataract is the leading cause of preventable blindness and visual impairment worldwide, particularly among elderly populations in low- and midd...
The impact of cataract surgery on tear film physiology: signs and symptoms, progression and treatment
The impact of cataract surgery on tear film physiology: signs and symptoms, progression and treatment
PurposeThis study aimed to revise data published in the literature on the effects of cataract surgery on tear film characteristics, in relation to personal clinical surgical experi...
Femtosecond laser LenSx–assisted phacoemulsification of mature intumescent cataract
Femtosecond laser LenSx–assisted phacoemulsification of mature intumescent cataract
Background: Phacoemulsification (phaco) of mature intumescent cataract is one of the most difficult to perform anterior eye surgeries. In mature intumescent cataract phaco, the mos...
Symmetrical femtosecond laser arc incision in correcting corneal astigmatism in cataract patients
Symmetrical femtosecond laser arc incision in correcting corneal astigmatism in cataract patients
AIM: To evaluate the effect of symmetrical arc incision correcting corneal astigmatism in femtosecond laser-assisted phacoemulsification (FLACS).
METHODS: This study enrolled patie...
Relationship Between Cataract Surgery and Mortality in Elderly Patients with Cataract: Nationwide Population-based Cohort Study in South Korea
Relationship Between Cataract Surgery and Mortality in Elderly Patients with Cataract: Nationwide Population-based Cohort Study in South Korea
Abstract
There were a few studies showing conflicting findings with regard to the relationship between cataract surgery and all-cause mortality. It is uncertain whether cat...
Manifestation of Corneal Wound Gap during Cataract Surgery: A Case Report
Manifestation of Corneal Wound Gap during Cataract Surgery: A Case Report
Significance: Intraoperative AS-OCT showed great advantages in the treatment of the patient with traumatic cataract and corneal perforation, and could be used to real-time monitor ...