Automatic detection of simulated motion blur in mammograms
View through CrossRef
PurposeTo use machine‐learning algorithms and blur measure (BM) operators to automatically detect motion blur in mammograms. Motion blur has been reported to reduce lesion detection performance and mask small abnormalities, resulting in failure to detect them until they reach more advanced stages. Automatic detection of blur could support the clinical decision‐making process during the mammography exam by allowing for an immediate retake, thereby preventing unnecessary expense, time, and patient anxiety.MethodsBlur was simulated mathematically to mimic the real blur seen in clinical practice. The blur point‐spread‐function (PSF) mask is generated by distributing pixel intensity of an image pixel moving under random motion within the range of blur effect (the maximum amount of tissue motion allowed). The random motion trajectory vector is generated on a super‐sampled image frame to accommodate smaller substeps; the vector was then sampled on a regular pixel grid using subpixel linear interpolation to generate the blur PSF mask. This randomly generated motion trajectory is constrained by several factors: the effects of variations in tissue elasticity, imaging exposure time, and size of blur effect (motion boundary in millimeters) were examined. The blur mask is convolved with a mammogram to create blur. Five motion blur magnitudes (0.1, 0.25, 0.5, 1.0, and 1.5 mm) were simulated on 244 and 434 mammograms from the INbreast and DDSM databases, respectively. Blur was quantified using nine BM operators for each mammogram and at each blur level. The mammograms were assigned to training (70%) and testing (30%) datasets to train three machine‐learning classifiers: Ensemble Bagged Trees, fine Gaussian SVM, and weighted KNN, to distinguish five levels of blurred from unblurred mammograms, using six‐way classification.ResultsFor the INbreast mammograms, the average classification accuracies were 87.7%, 85.7%, and 85.7% for Ensemble Bagged Trees, fine Gaussian SVM, and weighted KNN, respectively, and the average classification accuracies for DDSM were 93.5%, 93.6%, and 92.7% for Ensemble Bagged Trees, fine Gaussian SVM, and weighted KNN, respectively.ConclusionsPreliminary results show the potential to detect simulated blur automatically using those methods. Although limited work has been done to quantify the effects of motion blur on radiologists’ performance, there is evidence that motion blur might not be detected visually by a human observer and could negatively affect radiologists’ lesion detection performance. As of this date, no other study has investigated the ability of machine‐learning classifiers and BM operators to detect motion blur in mammograms.
Title: Automatic detection of simulated motion blur in mammograms
Description:
PurposeTo use machine‐learning algorithms and blur measure (BM) operators to automatically detect motion blur in mammograms.
Motion blur has been reported to reduce lesion detection performance and mask small abnormalities, resulting in failure to detect them until they reach more advanced stages.
Automatic detection of blur could support the clinical decision‐making process during the mammography exam by allowing for an immediate retake, thereby preventing unnecessary expense, time, and patient anxiety.
MethodsBlur was simulated mathematically to mimic the real blur seen in clinical practice.
The blur point‐spread‐function (PSF) mask is generated by distributing pixel intensity of an image pixel moving under random motion within the range of blur effect (the maximum amount of tissue motion allowed).
The random motion trajectory vector is generated on a super‐sampled image frame to accommodate smaller substeps; the vector was then sampled on a regular pixel grid using subpixel linear interpolation to generate the blur PSF mask.
This randomly generated motion trajectory is constrained by several factors: the effects of variations in tissue elasticity, imaging exposure time, and size of blur effect (motion boundary in millimeters) were examined.
The blur mask is convolved with a mammogram to create blur.
Five motion blur magnitudes (0.
1, 0.
25, 0.
5, 1.
0, and 1.
5 mm) were simulated on 244 and 434 mammograms from the INbreast and DDSM databases, respectively.
Blur was quantified using nine BM operators for each mammogram and at each blur level.
The mammograms were assigned to training (70%) and testing (30%) datasets to train three machine‐learning classifiers: Ensemble Bagged Trees, fine Gaussian SVM, and weighted KNN, to distinguish five levels of blurred from unblurred mammograms, using six‐way classification.
ResultsFor the INbreast mammograms, the average classification accuracies were 87.
7%, 85.
7%, and 85.
7% for Ensemble Bagged Trees, fine Gaussian SVM, and weighted KNN, respectively, and the average classification accuracies for DDSM were 93.
5%, 93.
6%, and 92.
7% for Ensemble Bagged Trees, fine Gaussian SVM, and weighted KNN, respectively.
ConclusionsPreliminary results show the potential to detect simulated blur automatically using those methods.
Although limited work has been done to quantify the effects of motion blur on radiologists’ performance, there is evidence that motion blur might not be detected visually by a human observer and could negatively affect radiologists’ lesion detection performance.
As of this date, no other study has investigated the ability of machine‐learning classifiers and BM operators to detect motion blur in mammograms.
Related Results
Technical feasibility of automated blur detection in digital mammography using convolutional neural network
Technical feasibility of automated blur detection in digital mammography using convolutional neural network
Abstract
Background
The presence of a blurred area, depending on its localization, in a mammogram can limit diagnostic accuracy. The goal of this st...
Research on Cross-Correlative Blur Length Estimation Algorithm in Motion Blur Image
Research on Cross-Correlative Blur Length Estimation Algorithm in Motion Blur Image
This paper proposes a motion blur length estimation method that is applied to motion blur image restoration. This method applies a cross-correlation algorithm to multi-frame motion...
Comparison of prospective and retrospective motion correction for Magnetic Resonance Imaging of the brain - Master's Thesis in Physics
Comparison of prospective and retrospective motion correction for Magnetic Resonance Imaging of the brain - Master's Thesis in Physics
Head motion is one of the most common sources of artefacts for Magnetic Resonance Imaging (MRI) of the brain. Especially children, being intimidated by the dimensions and the noise...
Categorizing Motion: Story-Based Categorizations
Categorizing Motion: Story-Based Categorizations
Our most primary goal is to provide a motion categorization for moving entities. A motion categorization that is related to how humans categorize motion, i.e., that is cognitive ...
Method of determining the distance to the object by analyzing its image blur
Method of determining the distance to the object by analyzing its image blur
In modern integrated monitoring systems and systems of automated control of technological processes there are several essential algorithms and procedures for obtaining primary info...
Flow parsing and biological motion
Flow parsing and biological motion
AbstractFlow parsing is a way to estimate the direction of scene-relative motion of independently moving objects during self-motion of the observer. So far, this has been tested fo...
Study on the partial discharge characteristics induced by the motion of cellulose particles in transformer oil
Study on the partial discharge characteristics induced by the motion of cellulose particles in transformer oil
AbstractCellulose particles present a significant concern within the oil‐paper insulation of transformers, posing potential risks to insulation performance. Under the influence of ...
Event‐by‐event respiratory motion correction for PET with 3D internal‐1D external motion correlation
Event‐by‐event respiratory motion correction for PET with 3D internal‐1D external motion correlation
Purpose:Respiratory motion during PET/CT imaging can cause substantial image blurring and underestimation of tracer concentration for both static and dynamic studies. In this study...