Search engine for discovering works of Art, research articles, and books related to Art and Culture
ShareThis
Javascript must be enabled to continue!

ESTIMASI NILAI DOSIS PADA PEMERIKSAAN CT SCAN KEPALA DEWASA BERDASARKAN NILAI CTDIvol DAN DLP

View through CrossRef
ABSTRACT   Background: Estimating the exact dose in CT Scan patients can be done by referring to an optimization index called DRL (Diagnostic Reference Level). Currently, the DRL for CT-Scans is expressed in CTDIvol (Computed Tomography Dose Index Volume) and DLP (Dose Length Product). The purpose of this study was to analyze the distribution of radiation dose values ​​on a CT scan of the adult head based on the CTDIvol and DLP values ​​referring to the IDRL reference from BAPETEN. Methods: This research is descriptive quantitative research with an observational approach. The Samples in this study were all patients who underwent Non-Contrast Head CT-Scan examinations at the radiology installation of Prof.Dr.I.G.N.G. Ngoerah Hospital, Denpasar for the period January-June 2022. Analysis of Results The estimation of radiation dose values ​​on CT Scan examinations of the adult head was based on the recapitulation of the distribution of CTDIvol and DLP values ​​obtained then compared with the IDRL 2021 values. Results: CTDIvol values ​​on non-Contrast head CT examination for all patients have varied values ​​with the lowest value of 23.28 mGy, the highest value of 62.7 mGy, an average of 47.58 mGy, and the 3rd quartile value of CTDIvol is 49.84 mGy. With the 75th percentile CTDIvol value of 49.84 mGy while the DLP value on the Head CT Examination varied in each patient with the lowest DLP value of 700.12 mGy*cm and the highest DLP value of 1597.19 mGy*cm and an average value of 1597.19 mGy*cm. DLP is 1057.95 mGy*Cm and the 3rd quartile value of DLP is 1133 mGy*cm. Conclusion: The CTDIvol and DLP values ​​for Non-Contrast Adult Head CT Scan examination at Prof.Dr.I.G.N.G. Ngoerah Hospital Denpasar are still below the reference from BAPETEN in IDRL 2021.
Title: ESTIMASI NILAI DOSIS PADA PEMERIKSAAN CT SCAN KEPALA DEWASA BERDASARKAN NILAI CTDIvol DAN DLP
Description:
ABSTRACT   Background: Estimating the exact dose in CT Scan patients can be done by referring to an optimization index called DRL (Diagnostic Reference Level).
Currently, the DRL for CT-Scans is expressed in CTDIvol (Computed Tomography Dose Index Volume) and DLP (Dose Length Product).
The purpose of this study was to analyze the distribution of radiation dose values ​​on a CT scan of the adult head based on the CTDIvol and DLP values ​​referring to the IDRL reference from BAPETEN.
Methods: This research is descriptive quantitative research with an observational approach.
The Samples in this study were all patients who underwent Non-Contrast Head CT-Scan examinations at the radiology installation of Prof.
Dr.
I.
G.
N.
G.
Ngoerah Hospital, Denpasar for the period January-June 2022.
Analysis of Results The estimation of radiation dose values ​​on CT Scan examinations of the adult head was based on the recapitulation of the distribution of CTDIvol and DLP values ​​obtained then compared with the IDRL 2021 values.
Results: CTDIvol values ​​on non-Contrast head CT examination for all patients have varied values ​​with the lowest value of 23.
28 mGy, the highest value of 62.
7 mGy, an average of 47.
58 mGy, and the 3rd quartile value of CTDIvol is 49.
84 mGy.
With the 75th percentile CTDIvol value of 49.
84 mGy while the DLP value on the Head CT Examination varied in each patient with the lowest DLP value of 700.
12 mGy*cm and the highest DLP value of 1597.
19 mGy*cm and an average value of 1597.
19 mGy*cm.
DLP is 1057.
95 mGy*Cm and the 3rd quartile value of DLP is 1133 mGy*cm.
Conclusion: The CTDIvol and DLP values ​​for Non-Contrast Adult Head CT Scan examination at Prof.
Dr.
I.
G.
N.
G.
Ngoerah Hospital Denpasar are still below the reference from BAPETEN in IDRL 2021.

Related Results

Penerapan Optimisasi Paparan Medik CT-Scan Melalui Tingkat Panduan Diagnostik di RSUP dr. Soeradji Tirtonegoro
Penerapan Optimisasi Paparan Medik CT-Scan Melalui Tingkat Panduan Diagnostik di RSUP dr. Soeradji Tirtonegoro
Latar belakang: Pemanfaatan CT-Scan di RSUP dr. Soeradji Tirtonegoro menunjukkan adanya peningkatan yang signifikan, pemanfaatan tersebut harus dilakukan pengawasan untuk menjamin ...
FAKTOR-FAKTOR YANG MEMPENGARUHI MORTALITAS PADA PASIEN DENGAN FRAKTUR COSTA: Literature Review
FAKTOR-FAKTOR YANG MEMPENGARUHI MORTALITAS PADA PASIEN DENGAN FRAKTUR COSTA: Literature Review
FAKTOR-FAKTOR YANG MEMPENGARUHI MORTALITAS PADA PASIEN DENGAN FRAKTUR COSTA: Literature  Review Anna Tri Wahyuni1), Masfuri2),  Liya Arista3)1,2,3 Fakultas Ilmu Keperawatan Univers...
OPTIMISASI PARAMETER SCAN PADA CT SCAN KEPALA NONKONTRAS DI INSTALASI RADIOLOGI RSUD PROF DR MARGONO SOEKARJO
OPTIMISASI PARAMETER SCAN PADA CT SCAN KEPALA NONKONTRAS DI INSTALASI RADIOLOGI RSUD PROF DR MARGONO SOEKARJO
Faktor eksposi yang sangat mempengaruhi dosis radiasi dan kualitas citra pada pemeriksaan CT Scan diantaranya tegangan (kV), kuat arus (mA) dan waktu pemindaian citra (s). Tegangan...
PENGGUNAAN PITCH PADA PEMERIKSAAN CT SCAN KEPALA TRAUMA DI RS PKU MUHAMMADIYAH YOGYAKARTA
PENGGUNAAN PITCH PADA PEMERIKSAAN CT SCAN KEPALA TRAUMA DI RS PKU MUHAMMADIYAH YOGYAKARTA
Pemeriksaan CT Scan kepala trauma memerlukan pengaturan parameter yang tepat, salah satunya adalah pitch. Pitch adalah rasio antara jarak pergerakan meja dan ketebalan irisan, yang...
PENERAPAN EDUKASI DAN KOMUNIKASI PASIEN MENGENAI DOSIS RADIASI SEBAGAI PENGEMBANGAN TINGKAT PANDUAN DIAGNOSTIK LOKAL
PENERAPAN EDUKASI DAN KOMUNIKASI PASIEN MENGENAI DOSIS RADIASI SEBAGAI PENGEMBANGAN TINGKAT PANDUAN DIAGNOSTIK LOKAL
Pemantauan dosis radiasi pada pasien merupakan aspek penting dalam upaya optimisasi proteksi radiasi. Penelitian ini bertujuan untuk menerapkan edukasi dan komunikasi pasien mengen...
Study the variations in radiation doses in different multi-slice CT scan machines
Study the variations in radiation doses in different multi-slice CT scan machines
Objectives: This study aimed to measure variations in dose output and patient’s effective dose across the different computed tomography scanners during standard CT examinations of ...
Analisis Dosis Efektif pada Pemeriksaan Computer Tomography (CT) Scan Abdomen di RSU Bali Jimbaran
Analisis Dosis Efektif pada Pemeriksaan Computer Tomography (CT) Scan Abdomen di RSU Bali Jimbaran
Telah dilakukan analisis dosis efektif dalam pemeriksaan CT Scan abdomen di RSU Bali Jimbaran. Penelitian dilakukan dengan menggunakan data sekunder hasil pemeriksaan CT Scan abdom...

Back to Top