On the development of a comprehensive MC simulation model for the Gamma Knife Perfexion radiosurgery unit

View through CrossRef
Abstract This work presents a comprehensive Monte Carlo (MC) simulation model for the Gamma Knife Perfexion (PFX) radiosurgery unit. Model-based dosimetry calculations were benchmarked in terms of relative dose profiles (RDPs) and output factors (OFs), against corresponding EBT2 measurements. To reduce the rather prolonged computational time associated with the comprehensive PFX model MC simulations, two approximations were explored and evaluated on the grounds of dosimetric accuracy. The first consists in directional biasing of the 60 Co photon emission while the second refers to the implementation of simplified source geometric models. The effect of the dose scoring volume dimensions in OF calculations accuracy was also explored. RDP calculations for the comprehensive PFX model were found to be in agreement with corresponding EBT2 measurements. Output factors of 0.819  ±  0.004 and 0.8941  ±  0.0013 were calculated for the 4 mm and 8 mm collimator, respectively, which agree, within uncertainties, with corresponding EBT2 measurements and published experimental data. Volume averaging was found to affect OF results by more than 0.3% for scoring volume radii greater than 0.5 mm and 1.4 mm for the 4 mm and 8 mm collimators, respectively. Directional biasing of photon emission resulted in a time efficiency gain factor of up to 210 with respect to the isotropic photon emission. Although no considerable effect on relative dose profiles was detected, directional biasing led to OF overestimations which were more pronounced for the 4 mm collimator and increased with decreasing emission cone half-angle, reaching up to 6% for a 5° angle. Implementation of simplified source models revealed that omitting the sources’ stainless steel capsule significantly affects both OF results and relative dose profiles, while the aluminum-based bushing did not exhibit considerable dosimetric effect. In conclusion, the results of this work suggest that any PFX simulation model should be benchmarked in terms of both RDP and OF results.
Title: On the development of a comprehensive MC simulation model for the Gamma Knife Perfexion radiosurgery unit
Description:
Abstract This work presents a comprehensive Monte Carlo (MC) simulation model for the Gamma Knife Perfexion (PFX) radiosurgery unit.
Model-based dosimetry calculations were benchmarked in terms of relative dose profiles (RDPs) and output factors (OFs), against corresponding EBT2 measurements.
To reduce the rather prolonged computational time associated with the comprehensive PFX model MC simulations, two approximations were explored and evaluated on the grounds of dosimetric accuracy.
The first consists in directional biasing of the 60 Co photon emission while the second refers to the implementation of simplified source geometric models.
The effect of the dose scoring volume dimensions in OF calculations accuracy was also explored.
RDP calculations for the comprehensive PFX model were found to be in agreement with corresponding EBT2 measurements.
Output factors of 0.
819  ±  0.
004 and 0.
8941  ±  0.
0013 were calculated for the 4 mm and 8 mm collimator, respectively, which agree, within uncertainties, with corresponding EBT2 measurements and published experimental data.
Volume averaging was found to affect OF results by more than 0.
3% for scoring volume radii greater than 0.
5 mm and 1.
4 mm for the 4 mm and 8 mm collimators, respectively.
Directional biasing of photon emission resulted in a time efficiency gain factor of up to 210 with respect to the isotropic photon emission.
Although no considerable effect on relative dose profiles was detected, directional biasing led to OF overestimations which were more pronounced for the 4 mm collimator and increased with decreasing emission cone half-angle, reaching up to 6% for a 5° angle.
Implementation of simplified source models revealed that omitting the sources’ stainless steel capsule significantly affects both OF results and relative dose profiles, while the aluminum-based bushing did not exhibit considerable dosimetric effect.
In conclusion, the results of this work suggest that any PFX simulation model should be benchmarked in terms of both RDP and OF results.

Related Results

On Proximal Relations in Transformation Semigroups Arising from Generalized Shifts
On Proximal Relations in Transformation Semigroups Arising from Generalized Shifts
For a finite discrete topological space $X$ with at least two elements, a nonempty set $\Gamma$, and a map $\varphi:\Gamma \to \Gamma$, $\sigma_{\varphi}:X^{\Gamma} \to X^{\Gamma}$...
P03.06 Cavernous malformations of the brainstem: radiosurgical management with gamma knife
P03.06 Cavernous malformations of the brainstem: radiosurgical management with gamma knife
Abstract BACKGROUND About 20% of intracranial cavernous malformations (CM) occur in the brainstem. The greatest risk of brainste...
An MR Imaging Study of Communicating Hydrocephalus After Gamma Knife Radiosurgery for Vestibular Schwannoma
An MR Imaging Study of Communicating Hydrocephalus After Gamma Knife Radiosurgery for Vestibular Schwannoma
Patients with vestibular schwannoma (VS) face a 3.7-23.7% risk developing hydrocephalus. Small-sized VS are associated with communicating or non-communicating hydrocephalus. Hydroc...
North Syrian Mortaria and Other Late Roman Personal and Utility Objects Bearing Inscriptions of Good Luck
North Syrian Mortaria and Other Late Roman Personal and Utility Objects Bearing Inscriptions of Good Luck
<span style="font-size: 11pt; color: black; font-family: 'Times New Roman','serif'">&Pi;&Eta;&Lambda;&Iota;&Nu;&Alpha; &Iota;&Gamma;&Delta...
AVM Radiosurgery: Where is the Volume Limit?
AVM Radiosurgery: Where is the Volume Limit?
To evaluate the role of radiosurgery for large AVM (nidus volume > 8 ml). 200 cerebral AVM patients treated by Gamma Knife radiosurgery constituted the study database. Of them, ...
OS8.3 Gamma knife radiosurgery for large meningiomas: 10 years of experience
OS8.3 Gamma knife radiosurgery for large meningiomas: 10 years of experience
Abstract BACKGROUND The use of gamma knife radiosurgery for the treatment of intracranial meningiomas has been established as an...
P27.32.B V-REX RANDOMIZED TRIAL
P27.32.B V-REX RANDOMIZED TRIAL
Abstract IMPORTANCE Current guidelines for treating small- to medium-sized vestibular schwannoma recommend either upfront radios...

Back to Top