Assessing lead curtains’ impact on radiation protection in coronary interventions

View through CrossRef
Abstract The objective of this investigation is to assess the impact of supplementary lead curtains on the reduction of radiation dose exposure to operators during coronary interventional procedures. Seven standard positions during coronary angiography (foot, right foot, head, left foot, left lateral, left head, and right lateral) were simulated on a standard anthropomorphic phantom with radial artery access. Measurements were taken at two different heights, 125 cm and 155 cm, and dosimeters were used to measured surface incident dose rates for the first and second operators, both with and without additional lead curtains at various positions. Each position was measured 20 times, and arithmetic means were computed. A t-test was utilised to compare dose rates with and without supplementary lead curtains, as well as dose rates with additional lead curtains at varying heights. The finding indicate that the dose rates of the first operator with supplementary lead curtains were not significantly lower compared to those without, except for the 125 cm head and left foot positions and the 155 cm head position with the additional lead curtain edge 10 cm below the umbilical level (t umbilical = 0.9, 0.4, 0.5, P > 0.05). The dose rates of the second operator with additional lead curtains were significantly lower than those without, with statistically significant differences (P < 0.05). The arithmetic mean dose rates for the first and second operators at each position were lowest when the upper edge of the additional lead curtain was situated 10 cm above the umbilical level. Employing supplementary lead curtains during coronary interventions effectively reduces radiation doses received by operators. The protective effect is enhanced when the additional lead curtain is closer to the irradiation field. Hence, it is recommended that additional curtains be employed judiciously, while ensuring that clinical procedures are not impeded, in order to effectively mitigate the radiation exposure of operators.
Title: Assessing lead curtains’ impact on radiation protection in coronary interventions
Description:
Abstract The objective of this investigation is to assess the impact of supplementary lead curtains on the reduction of radiation dose exposure to operators during coronary interventional procedures.
Seven standard positions during coronary angiography (foot, right foot, head, left foot, left lateral, left head, and right lateral) were simulated on a standard anthropomorphic phantom with radial artery access.
Measurements were taken at two different heights, 125 cm and 155 cm, and dosimeters were used to measured surface incident dose rates for the first and second operators, both with and without additional lead curtains at various positions.
Each position was measured 20 times, and arithmetic means were computed.
A t-test was utilised to compare dose rates with and without supplementary lead curtains, as well as dose rates with additional lead curtains at varying heights.
The finding indicate that the dose rates of the first operator with supplementary lead curtains were not significantly lower compared to those without, except for the 125 cm head and left foot positions and the 155 cm head position with the additional lead curtain edge 10 cm below the umbilical level (t umbilical = 0.
9, 0.
4, 0.
5, P > 0.
05).
The dose rates of the second operator with additional lead curtains were significantly lower than those without, with statistically significant differences (P < 0.
05).
The arithmetic mean dose rates for the first and second operators at each position were lowest when the upper edge of the additional lead curtain was situated 10 cm above the umbilical level.
Employing supplementary lead curtains during coronary interventions effectively reduces radiation doses received by operators.
The protective effect is enhanced when the additional lead curtain is closer to the irradiation field.
Hence, it is recommended that additional curtains be employed judiciously, while ensuring that clinical procedures are not impeded, in order to effectively mitigate the radiation exposure of operators.

Related Results

Digital Mental Health Landscaping in Low- and Middle-Income Countries 
Digital Mental Health Landscaping in Low- and Middle-Income Countries 
Introduction The aim of this project was to map the landscape of who is doing what and where in digital mental health, and to pr...
Early Onset of Coronary Subclavian Steal Syndrome: A Case Report and Literature Review
Early Onset of Coronary Subclavian Steal Syndrome: A Case Report and Literature Review
Abstract Introduction Coronary subclavian steal syndrome (CSSS) is a rare phenomenon that often goes undiagnosed and causes severe complications, including death. This report prese...
Magnetic resonance measurement of coronary blood flow
Magnetic resonance measurement of coronary blood flow
Magnetic resonance (MR) flow measurement in the coronary artery can be achieved with either a breath‐hold acquisition or a respiration‐triggered acquisition. MR measurements of car...
Ritelatur Review: Prosedur Pemeriksaan Coronary Angiogrhpy (Cag) Pada Kasus Acute Trasmular Myocardial Infraction
Ritelatur Review: Prosedur Pemeriksaan Coronary Angiogrhpy (Cag) Pada Kasus Acute Trasmular Myocardial Infraction
Literature review: coronary angiography examination procedure in cases of acute trasmular myocardial infarction. Background: Acute transmural myocardial infarction (ATMI) is a crit...
Thoracic radiotherapy in patients with lymphoma and restenosis after coronary stent placement
Thoracic radiotherapy in patients with lymphoma and restenosis after coronary stent placement
AbstractObjectives:The aim of this study was to assess the incidence of restenosis after coronary stenting in patients with lymphoma treated with thoracic radiation. Background: Pa...

Back to Top