Insufficient quality of public automated external defibrillator recordings in the greater Paris area, a descriptive study
View through CrossRef
BackgroundWestern countries report a significant increase in the proportion of patients who experience out-of-hospital cardiac arrests (OHCAs) and benefit from a public automated external defibrillator (pAED) before the arrival of rescue teams. However, recordings of devices recovered after resuscitation are of variable quality. Analysis of these data may inform decisions of whether to implement an internal defibrillator for survivors, and provide useful information about the performance of pAED algorithms and the actions of bystanders.ObjectiveTo investigate the quality of the information recorded by pAEDs during OHCAs in the Paris area.MethodspAED files used for some of the 8629 OHCAs that occurred in the greater Paris area between 1 January 2017 and 31 April 2019 on the day of the arrest were collected. The presence and accuracy of 23 factors required to interpret the recording was noted, including readability of the ECG, the presence of an impedance curve and the accuracy of the date and time. The recordings were analysed to assess the diagnostic and therapeutic performance of the pAEDs used.ResultsA total of 258 patients with an OHCA received assistance from a pAED, and 182 recordings were recovered. The pAEDs were made by 12 different manufacturers. Data extraction required eight different transmission modes and 16 software programmes; recordings were of highly heterogeneous quality. Two per cent of the recordings were of such poor quality that they were not interpretable. Among the 98% remaining, only 43% included a thoracic impedance curve, 34% the intensity of the shocks delivered and 8% the patient name. The date and time were accurate in 68% and 48% of recordings, respectively. The pAEDs had 87.6% (95% CI 83.7% to 91.0%) sensitivity and 99.5% (99.5% to 99.5%) specificity for defibrillating shockable rhythms (positive predictive value 98.2% (96.4% to 99.0%), negative predictive value 96.4% (95.3% to 96.8%)). The absence of important variables prevented the analysis of approximately half of the inappropriate decisions made by pAEDs.ConclusionCollection of pAED recordings is a major challenge. Their analysis is compromised by heterogeneity and poor quality (incomplete maintenance records, patient details and logs). AED recordings are currently the most relevant resource to track pAED performance and bystander practices. The quality of these recordings needs to improve.
Title: Insufficient quality of public automated external defibrillator recordings in the greater Paris area, a descriptive study
Description:
BackgroundWestern countries report a significant increase in the proportion of patients who experience out-of-hospital cardiac arrests (OHCAs) and benefit from a public automated external defibrillator (pAED) before the arrival of rescue teams.
However, recordings of devices recovered after resuscitation are of variable quality.
Analysis of these data may inform decisions of whether to implement an internal defibrillator for survivors, and provide useful information about the performance of pAED algorithms and the actions of bystanders.
ObjectiveTo investigate the quality of the information recorded by pAEDs during OHCAs in the Paris area.
MethodspAED files used for some of the 8629 OHCAs that occurred in the greater Paris area between 1 January 2017 and 31 April 2019 on the day of the arrest were collected.
The presence and accuracy of 23 factors required to interpret the recording was noted, including readability of the ECG, the presence of an impedance curve and the accuracy of the date and time.
The recordings were analysed to assess the diagnostic and therapeutic performance of the pAEDs used.
ResultsA total of 258 patients with an OHCA received assistance from a pAED, and 182 recordings were recovered.
The pAEDs were made by 12 different manufacturers.
Data extraction required eight different transmission modes and 16 software programmes; recordings were of highly heterogeneous quality.
Two per cent of the recordings were of such poor quality that they were not interpretable.
Among the 98% remaining, only 43% included a thoracic impedance curve, 34% the intensity of the shocks delivered and 8% the patient name.
The date and time were accurate in 68% and 48% of recordings, respectively.
The pAEDs had 87.
6% (95% CI 83.
7% to 91.
0%) sensitivity and 99.
5% (99.
5% to 99.
5%) specificity for defibrillating shockable rhythms (positive predictive value 98.
2% (96.
4% to 99.
0%), negative predictive value 96.
4% (95.
3% to 96.
8%)).
The absence of important variables prevented the analysis of approximately half of the inappropriate decisions made by pAEDs.
ConclusionCollection of pAED recordings is a major challenge.
Their analysis is compromised by heterogeneity and poor quality (incomplete maintenance records, patient details and logs).
AED recordings are currently the most relevant resource to track pAED performance and bystander practices.
The quality of these recordings needs to improve.
Related Results
Potable Water Sources, Household Hygiene, and Sanitation Practices in Ikpoba Okha LGA, Edo State: Implications for Public Health and Sustainable Water Management
Omoregie, Andrew Edosa.1 Omoregie Abieyuwa Peace2 Okoro, Enyinnaya Okoro.3
1 College of Medi
Potable Water Sources, Household Hygiene, and Sanitation Practices in Ikpoba Okha LGA, Edo State: Implications for Public Health and Sustainable Water Management
Omoregie, Andrew Edosa.1 Omoregie Abieyuwa Peace2 Okoro, Enyinnaya Okoro.3
1 College of Medi
BACKGROUND
Access to potable drinking water and sufficient sanitation continues to be an urgent global concern, particularly in developing regions where con...
Availability and use of Automated External Defibrillators in Karachi, Pakistan: a cross-sectional study
Availability and use of Automated External Defibrillators in Karachi, Pakistan: a cross-sectional study
Objective: To assess the availability and use of automated external defibrillators in various public and private establishments, and to assess knowledge, attitude and practices rel...
Exploring Large Language Models Integration in the Histopathologic Diagnosis of Skin Diseases: A Comparative Study
Exploring Large Language Models Integration in the Histopathologic Diagnosis of Skin Diseases: A Comparative Study
Abstract
Introduction
The exact manner in which large language models (LLMs) will be integrated into pathology is not yet fully comprehended. This study examines the accuracy, bene...
The effects of age on quality of life in implantable cardioverter defibrillator recipients
The effects of age on quality of life in implantable cardioverter defibrillator recipients
Background. The implantable cardioverter defibrillator shows superiority over conventional pharmacological therapy. The implantable cardioverter defibrillator has been implanted w...
If I Had Possession over Judgment Day: Augmenting Robert Johnson
If I Had Possession over Judgment Day: Augmenting Robert Johnson
augmentvb [ɔːgˈmɛnt]1. to make or become greater in number, amount, strength, etc.; increase2. Music: to increase (a major or perfect interval) by a semitone (Collins English Dicti...
Non-Recommended Publishing Lists: Strategies for Detecting Deceitful Journals
Non-Recommended Publishing Lists: Strategies for Detecting Deceitful Journals
Abstract
The rapid growth of open access publishing (OAP) has significantly improved the accessibility and dissemination of scientific knowledge. However, this expansion has also c...
Empty, Emptier, Emptiest (Leeg, Leger, Leegst)
Empty, Emptier, Emptiest (Leeg, Leger, Leegst)
Vacancy in the Dutch office market: Driving around any city in the Netherlands will tell you the same thing. ‘For rent’ signs appear on every corner of the street. The most importa...
Assessment of emergency physicians’ performance in identifying shockable rhythm in out-of-hospital cardiac arrest: an observational simulation study
Assessment of emergency physicians’ performance in identifying shockable rhythm in out-of-hospital cardiac arrest: an observational simulation study
BackgroundEmergency physicians can use a manual or an automated defibrillator to provide defibrillation of patients who had out-of-hospital cardiac arrest (OHCA). Performance of em...