Arthroscopic versus Open MTP Arthrodesis
View through CrossRef
Category:
Midfoot/Forefoot
Introduction/Purpose:
Hallux metatarsophalangeal (MTP) arthrodesis is a common procedure for painful conditions of the great toe. Dorsal plate fixation for MTP arthrodesis using locked plates produces good clinical outcomes and superior biomechanical strength to other techniques. However, arthroscopic fusion with new generation full thread compression screws is emerging as an alternative to open fusion. This method has been utilized clinically with good outcomes, but the biomechanical strength of arthroscopic MTP fusion fixation techniques is unknown. The purpose of this study was to compare low profile contoured locked plates to new generation full thread compression screws for first MTP fusion, in a biomechanical cadaver model. We hypothesize that there will be no significant difference in plantar gapping during cyclic loading, stiffness, or load-to-failure between the two groups,
Methods:
The first rays of eight matched pairs of fresh frozen cadaveric feet underwent dissection and DEXA scanning to measure bone mineral density (BMD). The “plate” group was prepared with cup-and-cone reamers, and fixation of the MTP joint with one compression screw and low profile dorsal locked plate. The matched pair “screws” group was prepared through a simulated arthroscopic technique, achieving fixation with two new generation full-thread compression screws, while preserving capsular attachments. Each specimen was loaded on the proximal phalanx in a cantilever fashion to 90 N at 3 Hz for a total of 250,000 cycles. Plantar MTP gap was recorded using a calibrated extensometer; Load-to-failure testing was performed for all specimens that endured cyclical loading; and stiffness was calculated from the final load-to-failure. Data was analyzed with a Student’s T-Test, with significance set at p<.05. Pearson Correlation coefficient (r) was calculated for stiffness and load-to-failure vs. BMD.
Results:
The plate group demonstrated significantly more plantar gapping during cyclic loading. There was no significant difference in stiffness, 31.6 N/mm (plates) and 51.7 N/mm (screws) (p=0.24) or load-to-failure, 198.6 N (plates) and 290.1 N (screws) (p =.07). Two of 8 screws-only specimens, and 3 of 8 locked plate specimens failed during cyclic loading. These early failures, and stiffness and load-to-failure were highly correlated to BMD for plates (r=0.85 and r=0.62, respectively) and screws only (r=0.82 and r=0.94, respectively). Maximum metatarsal head width measured on lateral view was strongly correlated with load-to-failure and stiffness for both groups (r > 0.7).
Conclusion:
Arthroscopic hallux MTP arthrodesis utilizing full thread compression screws has similar mean stiffness and load-to- failure compared to a low-profile locking plate, with significantly less plantar gapping. These data, combined with advantages of the arthroscopic preparation technique, support an increased role of arthroscopic fusion in lieu of more invasive open plating techniques. The two preload failures of the screws-only cohort occurred in specimens with the lowest BMD, potentially indicating a clinical contraindication with this technique. BMD and metatarsal width may aid in predicting early failure such that appropriate fixation construct and more conservative post-operative protocols might improve results for these patients.
Title: Arthroscopic versus Open MTP Arthrodesis
Description:
Category:
Midfoot/Forefoot
Introduction/Purpose:
Hallux metatarsophalangeal (MTP) arthrodesis is a common procedure for painful conditions of the great toe.
Dorsal plate fixation for MTP arthrodesis using locked plates produces good clinical outcomes and superior biomechanical strength to other techniques.
However, arthroscopic fusion with new generation full thread compression screws is emerging as an alternative to open fusion.
This method has been utilized clinically with good outcomes, but the biomechanical strength of arthroscopic MTP fusion fixation techniques is unknown.
The purpose of this study was to compare low profile contoured locked plates to new generation full thread compression screws for first MTP fusion, in a biomechanical cadaver model.
We hypothesize that there will be no significant difference in plantar gapping during cyclic loading, stiffness, or load-to-failure between the two groups,
Methods:
The first rays of eight matched pairs of fresh frozen cadaveric feet underwent dissection and DEXA scanning to measure bone mineral density (BMD).
The “plate” group was prepared with cup-and-cone reamers, and fixation of the MTP joint with one compression screw and low profile dorsal locked plate.
The matched pair “screws” group was prepared through a simulated arthroscopic technique, achieving fixation with two new generation full-thread compression screws, while preserving capsular attachments.
Each specimen was loaded on the proximal phalanx in a cantilever fashion to 90 N at 3 Hz for a total of 250,000 cycles.
Plantar MTP gap was recorded using a calibrated extensometer; Load-to-failure testing was performed for all specimens that endured cyclical loading; and stiffness was calculated from the final load-to-failure.
Data was analyzed with a Student’s T-Test, with significance set at p<.
05.
Pearson Correlation coefficient (r) was calculated for stiffness and load-to-failure vs.
BMD.
Results:
The plate group demonstrated significantly more plantar gapping during cyclic loading.
There was no significant difference in stiffness, 31.
6 N/mm (plates) and 51.
7 N/mm (screws) (p=0.
24) or load-to-failure, 198.
6 N (plates) and 290.
1 N (screws) (p =.
07).
Two of 8 screws-only specimens, and 3 of 8 locked plate specimens failed during cyclic loading.
These early failures, and stiffness and load-to-failure were highly correlated to BMD for plates (r=0.
85 and r=0.
62, respectively) and screws only (r=0.
82 and r=0.
94, respectively).
Maximum metatarsal head width measured on lateral view was strongly correlated with load-to-failure and stiffness for both groups (r > 0.
7).
Conclusion:
Arthroscopic hallux MTP arthrodesis utilizing full thread compression screws has similar mean stiffness and load-to- failure compared to a low-profile locking plate, with significantly less plantar gapping.
These data, combined with advantages of the arthroscopic preparation technique, support an increased role of arthroscopic fusion in lieu of more invasive open plating techniques.
The two preload failures of the screws-only cohort occurred in specimens with the lowest BMD, potentially indicating a clinical contraindication with this technique.
BMD and metatarsal width may aid in predicting early failure such that appropriate fixation construct and more conservative post-operative protocols might improve results for these patients.
Related Results
A Direct Referencing Method of the Tibial Plateau for the Posterior Tibial Slope in Medial Unicompartmental Knee Arthroplasty
A Direct Referencing Method of the Tibial Plateau for the Posterior Tibial Slope in Medial Unicompartmental Knee Arthroplasty
Abstract
Purpose
There is no consensus on intraoperative references for the posterior tibial slope (PTS) in medial unicompartmental knee arthroplasty (UKA). An arthroscopi...
A direct referencing method of the tibial plateau for the posterior tibial slope in medial unicompartmental knee arthroplasty
A direct referencing method of the tibial plateau for the posterior tibial slope in medial unicompartmental knee arthroplasty
Abstract
Purpose
There is no consensus on intraoperative references for the posterior tibial slope (PTS) in medial unicompartmental knee arthroplast...
Arthroscopic Debridement for First Metatarsophalangeal Joint Arthrodesis With a 2‐ Versus 3‐Portal Technique: A Cadaveric Study
Arthroscopic Debridement for First Metatarsophalangeal Joint Arthrodesis With a 2‐ Versus 3‐Portal Technique: A Cadaveric Study
Purpose
To evaluate the safety and efficacy of arthroscopic debridement for arthrodesis of the first metatarsophalangeal (MTP) joint using a 2‐portal technique ...
Prevalence and distribution of sesamoid bones and accessory ossicles of the foot as determined by digital tomosynthesis
Prevalence and distribution of sesamoid bones and accessory ossicles of the foot as determined by digital tomosynthesis
We previously reported that digital tomosynthesis (DTS) is a reliable tool for evaluating bony structures of the hand. The current study aimed to identify the prevalence and distri...
Arthroscopic ankle surgery in people with haemophilia
Arthroscopic ankle surgery in people with haemophilia
AbstractIntroductionPeople with haemophilia (PWH) not administered primary haematological prophylaxis since childhood, that is, those treated haematologically on demand or not trea...
Arthroscopic Subtalar Joint Arthrodesis: Topical Review
Arthroscopic Subtalar Joint Arthrodesis: Topical Review
Background:
Advances in foot and ankle arthroscopy over the last 2 decades have allowed for subtalar joint arthrodesis to be performed arthroscopically. The pot...
Outcome Of Arthroscopic Vs Mini Open Subacromial Decompression In Treating Shoulder Impingement Syndrome
Outcome Of Arthroscopic Vs Mini Open Subacromial Decompression In Treating Shoulder Impingement Syndrome
Introduction: Shoulder impingement syndrome is a common diagnosis leading to shoulder disabilities. Subacromial decompression surgery either arthroscopic or mini open is indicated ...
The Need for Aggressive Correction of Hypocalcemia During Traumatic Massive Transfusion Protocol
The Need for Aggressive Correction of Hypocalcemia During Traumatic Massive Transfusion Protocol
Introduction
Hypocalcemia is a common side effect of massive transfusion protocols (MTPs) after traumatic injury. The relationship between hypocalcemia severity...