Diagnostic Performance of the PalmScan VF2000 Virtual Reality Visual Field Analyzer for Identification and Classification of Glaucoma
View through CrossRef
Purpose: To evaluate the diagnostic test properties of the Palm Scan VF2000® Virtual Reality Visual Field Analyzer for diagnosis and classification of the severity of glaucoma.
Methods: This study was a prospective cross-sectional analysis of 166 eyes from 97 participants. All of them were examined by the Humphrey® Field Analyzer (used as the gold standard) and the Palm Scan VF 2000® Virtual Reality Visual Field Analyzer on the same day by the same examiner. We estimated the kappa statistic (including 95% confidence interval [CI]) as a measure of agreement between these two methods. The diagnostic test properties were assessed using sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV).
Results: The sensitivity, specificity, PPV, and NPV for the Virtual Reality Visual Field Analyzer for the classification of individuals as glaucoma/non-glaucoma was 100%. The general agreement for the classification of glaucoma between these two instruments was 0.63 (95% CI: 0.56–0.78). The agreement for mild glaucoma was 0.76 (95% CI: 0.61–0.92), for moderate glaucoma was 0.37 (0.14–0.60), and for severe glaucoma was 0.70 (95% CI: 0.55–0.85). About 28% of moderate glaucoma cases were misclassified as mild and 17% were misclassified as severe by the virtual reality visual field analyzer. Furthermore, 20% of severe cases were misclassified as moderate by this instrument.
Conclusion: The instrument is 100% sensitive and specific in detection of glaucoma. However, among patients with glaucoma, there is a relatively high proportion of misclassification of severity of glaucoma. Thus, although useful for screening of glaucoma, it cannot replace the Humphrey® Field Analyzer for the clinical management in its current form.
Title: Diagnostic Performance of the PalmScan VF2000 Virtual Reality Visual Field Analyzer for Identification and Classification of Glaucoma
Description:
Purpose: To evaluate the diagnostic test properties of the Palm Scan VF2000® Virtual Reality Visual Field Analyzer for diagnosis and classification of the severity of glaucoma.
Methods: This study was a prospective cross-sectional analysis of 166 eyes from 97 participants.
All of them were examined by the Humphrey® Field Analyzer (used as the gold standard) and the Palm Scan VF 2000® Virtual Reality Visual Field Analyzer on the same day by the same examiner.
We estimated the kappa statistic (including 95% confidence interval [CI]) as a measure of agreement between these two methods.
The diagnostic test properties were assessed using sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV).
Results: The sensitivity, specificity, PPV, and NPV for the Virtual Reality Visual Field Analyzer for the classification of individuals as glaucoma/non-glaucoma was 100%.
The general agreement for the classification of glaucoma between these two instruments was 0.
63 (95% CI: 0.
56–0.
78).
The agreement for mild glaucoma was 0.
76 (95% CI: 0.
61–0.
92), for moderate glaucoma was 0.
37 (0.
14–0.
60), and for severe glaucoma was 0.
70 (95% CI: 0.
55–0.
85).
About 28% of moderate glaucoma cases were misclassified as mild and 17% were misclassified as severe by the virtual reality visual field analyzer.
Furthermore, 20% of severe cases were misclassified as moderate by this instrument.
Conclusion: The instrument is 100% sensitive and specific in detection of glaucoma.
However, among patients with glaucoma, there is a relatively high proportion of misclassification of severity of glaucoma.
Thus, although useful for screening of glaucoma, it cannot replace the Humphrey® Field Analyzer for the clinical management in its current form.
Related Results
Childhood glaucoma profile in a Southwestern Ethiopia tertiary care center: a retrospective study
Childhood glaucoma profile in a Southwestern Ethiopia tertiary care center: a retrospective study
Abstract
Background
Childhood glaucoma is a major cause of childhood blindness worldwide. The profile of childhood glaucoma has not been well characterized in sub-Saharan ...
Profile of Newly Referred Glaucoma Patients to the Largest Tertiary Eye Care Hospital in Saudi Arabia
Profile of Newly Referred Glaucoma Patients to the Largest Tertiary Eye Care Hospital in Saudi Arabia
Abstract
Purpose: To report the profile of newly referred glaucoma patients to the largest tertiary eye care hospital in Saudi Arabia.
Patients and Methods: Medical record...
Proportion of Glaucoma among Voluntary People Coming for Glaucoma Screening Program at Jimma University Department of Ophthalmology, Jimma, Ethiopia
Proportion of Glaucoma among Voluntary People Coming for Glaucoma Screening Program at Jimma University Department of Ophthalmology, Jimma, Ethiopia
BACKGROUND: Glaucoma is the second leading cause of blindness worldwide. Glaucoma screening has been controversial, and much of the controversy has centered on the value of mass or...
Trend of glaucoma internal filtration surgeries in a tertiary hospital in China
Trend of glaucoma internal filtration surgeries in a tertiary hospital in China
AIM: To evaluate the trend of glaucoma internal filtration surgeries for inpatients between 2015 and 2021 at the Eye Hospital of Wenzhou Medical University.
METHODS: A review of th...
Profile of pediatric glaucoma patients in Shanghai Eye, Ear, Nose and Throat Hospital
Profile of pediatric glaucoma patients in Shanghai Eye, Ear, Nose and Throat Hospital
BackgroundThe extent of our knowledge of the epidemiology of pediatric glaucoma in China is limited. To better characterize the epidemiology of pediatric glaucoma in eastern China,...
Comparing cybersickness in virtual reality and mixed reality head-mounted displays
Comparing cybersickness in virtual reality and mixed reality head-mounted displays
Introduction: Defence Research and Development Canada is developing guidance on the use of Mixed Reality head-mounted displays for naval operations in the Royal Canadian Navy. Virt...