Testing Spirometers: Are the Standard Curves of the American Thoracic Society Sufficient?
View through CrossRef
BACKGROUND:
The performance of spirometers is often measured only under ideal conditions, with a mechanical simulator reproducing the expiratory standard American Thoracic Society (ATS) curves generated by a computer. Studies have questioned the value of these results in real-life conditions. The aim of this study was to evaluate the accuracy and precision of 5 office spirometers with a flow-volume simulator using the ATS curves and using flow-volume curves obtained from patients.
METHODS:
We measured the FVC, peak expiratory flow, and FEV
1
by simulating different dynamic waveforms applied by a computer-driven syringe, the Hans Rudolph flow-volume simulator. In addition to testing standard curves recommended by the ATS, we also tested curves obtained with subjects.
RESULTS:
The precision of the office spirometers was good and comparable using the standard ATS curves. One device presented the best performances in terms of accuracy and precision according to the ATS recommendations, but we observed significant biases in all devices with Bland-Altman analysis, particularly with the curves obtained from subjects with severe COPD.
CONCLUSIONS:
The global quality of most spirometers makes them acceptable for the detection of pulmonary diseases. However, we demonstrated accuracy issues not shown by the standard testing procedure. We propose to improve the testing of spirometers by implementing more realistic flow-volume curves and to refine the analysis of the results.
Title: Testing Spirometers: Are the Standard Curves of the American Thoracic Society Sufficient?
Description:
BACKGROUND:
The performance of spirometers is often measured only under ideal conditions, with a mechanical simulator reproducing the expiratory standard American Thoracic Society (ATS) curves generated by a computer.
Studies have questioned the value of these results in real-life conditions.
The aim of this study was to evaluate the accuracy and precision of 5 office spirometers with a flow-volume simulator using the ATS curves and using flow-volume curves obtained from patients.
METHODS:
We measured the FVC, peak expiratory flow, and FEV
1
by simulating different dynamic waveforms applied by a computer-driven syringe, the Hans Rudolph flow-volume simulator.
In addition to testing standard curves recommended by the ATS, we also tested curves obtained with subjects.
RESULTS:
The precision of the office spirometers was good and comparable using the standard ATS curves.
One device presented the best performances in terms of accuracy and precision according to the ATS recommendations, but we observed significant biases in all devices with Bland-Altman analysis, particularly with the curves obtained from subjects with severe COPD.
CONCLUSIONS:
The global quality of most spirometers makes them acceptable for the detection of pulmonary diseases.
However, we demonstrated accuracy issues not shown by the standard testing procedure.
We propose to improve the testing of spirometers by implementing more realistic flow-volume curves and to refine the analysis of the results.
Related Results
Blunt Chest Trauma and Chylothorax: A Systematic Review
Blunt Chest Trauma and Chylothorax: A Systematic Review
Abstract
Introduction: Although traumatic chylothorax is predominantly associated with penetrating injuries, instances following blunt trauma, as a rare and challenging condition, ...
Presentation and Management of Cervical Thoracic Duct Cyst: A Systematic Review of the Literature
Presentation and Management of Cervical Thoracic Duct Cyst: A Systematic Review of the Literature
Abstract
Introduction
Thoracic duct cysts are an uncommon phenomenon, especially within the cervical region. Due to its limited reported cases, very little is known about its etiol...
Phrenic Nerve Block for Management of Post-Thoracic Outlet Decompression Cough: A Case Report and Literature Review
Phrenic Nerve Block for Management of Post-Thoracic Outlet Decompression Cough: A Case Report and Literature Review
Abstract
Introduction
Thoracic outlet syndrome is a group of disorders arising from compressive forces on the neurovascular bundle in that region due to different etiologies. This...
Divergent Conceptualizations and Management Strategies for Neurogenic Thoracic Outlet Syndrome: A Qualitative Multispecialty Study
Divergent Conceptualizations and Management Strategies for Neurogenic Thoracic Outlet Syndrome: A Qualitative Multispecialty Study
Abstract
Background
Neurogenic thoracic outlet syndrome (nTOS) is the most prevalent subtype of thoracic outlet syndrome and remains one of the most controversial conditions in per...
Sagittal and transversal plane deformity in thoracic scoliosis
Sagittal and transversal plane deformity in thoracic scoliosis
The aim of the study was to assess the sagittal and transversal plane deformity of the spine in thoracic scoliosis by the mean of 3-D radiographic analysis. 46 patients admitted fo...
Why Are Some Intervertebral Discs More Prone to Degeneration?
Why Are Some Intervertebral Discs More Prone to Degeneration?
Study Design.
Prospective observational study.
Objective.
To determine the prevalence of isolated thoracic degeneration on...
SOFTWARE TESTING TECHNIQUES AND PRINCIPLES
SOFTWARE TESTING TECHNIQUES AND PRINCIPLES
This paper describes Software testing, need for software testing, Software testing goals and principles. Further it describe about different Software testing techniques and differe...
Differential Diagnosis of Neurogenic Thoracic Outlet Syndrome: A Review
Differential Diagnosis of Neurogenic Thoracic Outlet Syndrome: A Review
Abstract
Thoracic outlet syndrome (TOS) is a complex and often overlooked condition caused by the compression of neurovascular structures as they pass through the thoracic outlet. ...