Bent Telescopic Rods in Patients With Osteogenesis Imperfecta
View through CrossRef
Background:
Telescopic rods require alignment of 2 rods to enable lengthening. A telescopic rod converts functionally into a solid rod if either rod bends, preventing proper engagement. Our goal was to characterize implant bending as a mode of failure of telescopic rods used in the treatment of osteogenesis imperfecta in children.
Methods:
We conducted a retrospective review of our osteogenesis imperfecta database for patients treated with intramedullary telescopic rods at our institution from 1992 through 2010 and identified 12 patients with bent rods. The 6 boys and 6 girls had an average age at the time of initial surgery of 3.1 years (range, 1.8 to 8.3 y) and a total of 51 telescoping rods. Clinic notes, operative reports, and radiographs were reviewed. The rods were analyzed for amount of lengthening, characteristics of bending, presence of cut out, or disengagement from an anchor point. Bends in the rods were characterized by their location on the implant component. The bent and straight rods were compared. Data were analyzed with the Mann-Whitney test (statistical significance set at P≤0.05).
Results:
Of the 51 telescoping rods, 17 constructs (33%) bent. The average interval between surgery and rod bending was 4.0 years (range, 0.9 to 8.2 y). Before bending, 11 of 17 telescoping rods had routine follow-up radiographs for review. In 10 of the rods, bending was present when early signs of rod failure were first detected. Rod bending did not seem to be related to rod size. There was no area on the rod itself that seemed more susceptible to bending.
Conclusions:
Rod bending can be an early sign of impending rod failure. When rod bending is first noted, it may predispose the rod to other subsequent failures such as loss of proximal and distal fixation and cut out. Rod bending should be viewed as an indicator for closer monitoring of the patient and discussions regarding future need for rod exchange.
Level of Evidence:
Level III—retrospective review.
Ovid Technologies (Wolters Kluwer Health)
Title: Bent Telescopic Rods in Patients With Osteogenesis Imperfecta
Description:
Background:
Telescopic rods require alignment of 2 rods to enable lengthening.
A telescopic rod converts functionally into a solid rod if either rod bends, preventing proper engagement.
Our goal was to characterize implant bending as a mode of failure of telescopic rods used in the treatment of osteogenesis imperfecta in children.
Methods:
We conducted a retrospective review of our osteogenesis imperfecta database for patients treated with intramedullary telescopic rods at our institution from 1992 through 2010 and identified 12 patients with bent rods.
The 6 boys and 6 girls had an average age at the time of initial surgery of 3.
1 years (range, 1.
8 to 8.
3 y) and a total of 51 telescoping rods.
Clinic notes, operative reports, and radiographs were reviewed.
The rods were analyzed for amount of lengthening, characteristics of bending, presence of cut out, or disengagement from an anchor point.
Bends in the rods were characterized by their location on the implant component.
The bent and straight rods were compared.
Data were analyzed with the Mann-Whitney test (statistical significance set at P≤0.
05).
Results:
Of the 51 telescoping rods, 17 constructs (33%) bent.
The average interval between surgery and rod bending was 4.
0 years (range, 0.
9 to 8.
2 y).
Before bending, 11 of 17 telescoping rods had routine follow-up radiographs for review.
In 10 of the rods, bending was present when early signs of rod failure were first detected.
Rod bending did not seem to be related to rod size.
There was no area on the rod itself that seemed more susceptible to bending.
Conclusions:
Rod bending can be an early sign of impending rod failure.
When rod bending is first noted, it may predispose the rod to other subsequent failures such as loss of proximal and distal fixation and cut out.
Rod bending should be viewed as an indicator for closer monitoring of the patient and discussions regarding future need for rod exchange.
Level of Evidence:
Level III—retrospective review.
Related Results
Treatment of Osteogenesis Imperfecta with Corkscrew Tipped Telescopic Nail
Treatment of Osteogenesis Imperfecta with Corkscrew Tipped Telescopic Nail
Introduction and Aim: Osteogenesis imperfecta (OI) cause health problems in all tissues contain type-1 collagen. The patient is complaining from easily fragile bones due to loss en...
Corkscrew Tipped Telescopic Nail for Treatment of Osteogenesis Imperfecta: A clinical Experience
Corkscrew Tipped Telescopic Nail for Treatment of Osteogenesis Imperfecta: A clinical Experience
Introduction and aim: Osteogenesis imperfecta (OI) cause health problems in all tissues contain type-1 collagen. The patient is complaining from easily fragile bones due to loss en...
Dynamic Simulation Analysis of Truck Crane Based on ADAMS
Dynamic Simulation Analysis of Truck Crane Based on ADAMS
Background:
With a reduction in the weight of telescopic boom crane and an increase in its
lifting capacity, the contradiction between its speed, stability and safety becomes more ...
HRQoL Evaluation of Pediatric Osteogenesis Imperfecta with Zoledronic Acid Therapy
HRQoL Evaluation of Pediatric Osteogenesis Imperfecta with Zoledronic Acid Therapy
Background: Zoledronic acid as bisphosphonates could increase bone mineral density (BMD), which in osteogenesis imperfecta will reduce clinical manifestations. Pediatric patients w...
Analysis of phenotypic manifestations of connective tissue dysplasia in adult patients with osteogenesis imperfecta and development of a diagnostic algorithm
Analysis of phenotypic manifestations of connective tissue dysplasia in adult patients with osteogenesis imperfecta and development of a diagnostic algorithm
Connective tissue dysplasia (CTD) is a genetically determined disorder of the development of connective tissue in the embryonic and postnatal periods, which is characterized by def...
Total knee replacement in Osteogenesis Imperfecta: a case report and review of the literature
Total knee replacement in Osteogenesis Imperfecta: a case report and review of the literature
Abstract
Background
A review of the literature revealed that only 9 total knee replacements were performed on patients with osteogenesis imperfecta ...
New 3D Cone Beam CT Imaging Parameters to Assist the Dentist in Treating Patients with Osteogenesis Imperfecta
New 3D Cone Beam CT Imaging Parameters to Assist the Dentist in Treating Patients with Osteogenesis Imperfecta
(1) Background: The aim of the work is to identify some imaging parameters in osteogenesis imperfecta to assist the dentist in the diagnosis, planning, and orthodontic treatment of...
The importance of preoperative nasal endoscopy in patients with sinonasal disease
The importance of preoperative nasal endoscopy in patients with sinonasal disease
AbstractFunctional endoscopic sinus surgery (FESS) has popularized the use of telescopes in sinus surgery and has emphasized the importance of nasal endoscopy and computed tomograp...