Cell-cultured PDMS vascular model to allow placement of implant devices

View through CrossRef
AbstractBackgroundA stent maintains normal blood flow by expanding a stenotic artery from the inside. Stents currently in clinical use are mechanically sub-optimized and, therefore, exert excessive mechanical stimuli on the vascular wall. This can lead to arterial inflammation and potentially in-stent restenosis. To optimize the force exerted by a stent on the vascular wall, we should understand, reproduce, and analyze in detail the mechanical field to which the stented vessel is exposed. However, to date, noin vitromodel has been established that can adequately evaluate the mechanical effects of a stent on the vascular wall and be constructed without relying on a laboratory’s know-how. This study aims to develop a three-dimensionalin vitrovascular model that allows stent placement and to provide the details of its construction method.MethodsPolydimethylsiloxane (PDMS) was used to mimic the adventitial structure of arteries. Human carotid artery endothelial cells (HCtAECs) were then seeded on the luminal surface and cultured for 24 h to form a confluent monolayer (intima). The constructed model was installed in the flow-exposure culturing system, and hemodynamic stimulation (two types of shear stress (SS); 0.5 Pa and 2.3 Pa) was applied to the HCtAECs inside to reproduce the physiological state of blood vessels. In addition, a self-expanding stent was placed in the model during perfusion culture.ResultsWe examined the performance of the developed model based on quantitative evaluations of endothelial morphology in response to SS and in-stent endothelialization. Exposure to SS for 24 and 48 h caused endothelial orientation and elongation in the direction of flow, confirming the physiological responses of blood vessels. We were also able to quantitatively evaluate endothelial migration (endothelialization) to the stent.DiscussionOur model can reproduce the mechanical field inside the stented blood vessel, including hemodynamic stimuli. Based on the results obtained, the developed cell-cultured vascular model has sufficient performance to be used not only for quantitative evaluation of in-stent endothelialization but also for optimization of the mechanical fields in the stented vessel.
Title: Cell-cultured PDMS vascular model to allow placement of implant devices
Description:
AbstractBackgroundA stent maintains normal blood flow by expanding a stenotic artery from the inside.
Stents currently in clinical use are mechanically sub-optimized and, therefore, exert excessive mechanical stimuli on the vascular wall.
This can lead to arterial inflammation and potentially in-stent restenosis.
To optimize the force exerted by a stent on the vascular wall, we should understand, reproduce, and analyze in detail the mechanical field to which the stented vessel is exposed.
However, to date, noin vitromodel has been established that can adequately evaluate the mechanical effects of a stent on the vascular wall and be constructed without relying on a laboratory’s know-how.
This study aims to develop a three-dimensionalin vitrovascular model that allows stent placement and to provide the details of its construction method.
MethodsPolydimethylsiloxane (PDMS) was used to mimic the adventitial structure of arteries.
Human carotid artery endothelial cells (HCtAECs) were then seeded on the luminal surface and cultured for 24 h to form a confluent monolayer (intima).
The constructed model was installed in the flow-exposure culturing system, and hemodynamic stimulation (two types of shear stress (SS); 0.
5 Pa and 2.
3 Pa) was applied to the HCtAECs inside to reproduce the physiological state of blood vessels.
In addition, a self-expanding stent was placed in the model during perfusion culture.
ResultsWe examined the performance of the developed model based on quantitative evaluations of endothelial morphology in response to SS and in-stent endothelialization.
Exposure to SS for 24 and 48 h caused endothelial orientation and elongation in the direction of flow, confirming the physiological responses of blood vessels.
We were also able to quantitatively evaluate endothelial migration (endothelialization) to the stent.
DiscussionOur model can reproduce the mechanical field inside the stented blood vessel, including hemodynamic stimuli.
Based on the results obtained, the developed cell-cultured vascular model has sufficient performance to be used not only for quantitative evaluation of in-stent endothelialization but also for optimization of the mechanical fields in the stented vessel.

Related Results

MARS-seq2.0: an experimental and analytical pipeline for indexed sorting combined with single-cell RNA sequencing v1
MARS-seq2.0: an experimental and analytical pipeline for indexed sorting combined with single-cell RNA sequencing v1
Human tissues comprise trillions of cells that populate a complex space of molecular phenotypes and functions and that vary in abundance by 4–9 orders of magnitude. Relying solely ...
In Vitro Evaluation of Zinc Oxide Tetrapods as a New Material Component for Glaucoma Implants
In Vitro Evaluation of Zinc Oxide Tetrapods as a New Material Component for Glaucoma Implants
In our previous study we were able to show that zinc oxide (ZnO) tetrapods inhibit wound healing processes. Therefore, the aim of this study was to test the antiproliferative effec...
Evaluation of hybrid implant in posterior edentulous maxilla with inadequate bone
Evaluation of hybrid implant in posterior edentulous maxilla with inadequate bone
Aim of this study is to investigate.Hybrid implant has got satisfactory results in our previous study in various edentulous areas with inadequate bone for endosseous implant placem...
Cochlear implant material effects on inflammatory cell function and foreign body response
Cochlear implant material effects on inflammatory cell function and foreign body response
AbstractObjectivesThe objectives of this study were to assess the effects of cochlear implant (CI) biomaterials on the function of macrophages and fibroblasts, two key mediators of...
Wearable porous PDMS layer of high moisture permeability for skin trouble reduction
Wearable porous PDMS layer of high moisture permeability for skin trouble reduction
AbstractThe present research proposes the present porous polydimethylsiloxane (PDMS) layer for the skin trouble reduced daily life skin attachable devices. The present research pro...

Back to Top