Javascript must be enabled to continue!
In vitro characterization of Lagrangian fluid transport downstream of a dysfunctional bileaflet mechanical aortic valve
View through CrossRef
This experimental study aims to explore the Lagrangian nature of fluid transport downstream of a bileaflet mechanical aortic valve under different malfunction scenarios that might be encountered clinically. Time-resolved planar particle image velocimetry measurements are performed to extract instantaneous velocity fields downstream of the bileaflet mechanical valve implanted in an elastic aortic model. The results show an increase in particle residence time with the severity of malfunction. This is attributed to the expansion of the recirculation regions downstream of the valve. The time-evolution of Lagrangian coherent structures over one cardiac cycle (using finite-time Lyapunov exponent fields) shows the effect of valve dysfunction on the material transport and its barriers inside the aorta. The unbalanced flow through the dysfunctional leaflets leads to a significant redistribution of the LCS, thus the fluid transport along the ascending aorta. Moreover, a new technique for the evaluation of the highest accumulated shear stresses is applied along the Lagrangian trajectory of particles being released from the extracted Lagrangian coherent structures where the highest stretching occurs. Finally, the induced non-laminar flow behavior by the valve dysfunction is analyzed using the time-frequency spectra of velocity signals at selected points in the ascending aorta.
Title: In vitro characterization of Lagrangian fluid transport downstream of a dysfunctional bileaflet mechanical aortic valve
Description:
This experimental study aims to explore the Lagrangian nature of fluid transport downstream of a bileaflet mechanical aortic valve under different malfunction scenarios that might be encountered clinically.
Time-resolved planar particle image velocimetry measurements are performed to extract instantaneous velocity fields downstream of the bileaflet mechanical valve implanted in an elastic aortic model.
The results show an increase in particle residence time with the severity of malfunction.
This is attributed to the expansion of the recirculation regions downstream of the valve.
The time-evolution of Lagrangian coherent structures over one cardiac cycle (using finite-time Lyapunov exponent fields) shows the effect of valve dysfunction on the material transport and its barriers inside the aorta.
The unbalanced flow through the dysfunctional leaflets leads to a significant redistribution of the LCS, thus the fluid transport along the ascending aorta.
Moreover, a new technique for the evaluation of the highest accumulated shear stresses is applied along the Lagrangian trajectory of particles being released from the extracted Lagrangian coherent structures where the highest stretching occurs.
Finally, the induced non-laminar flow behavior by the valve dysfunction is analyzed using the time-frequency spectra of velocity signals at selected points in the ascending aorta.
Related Results
Experimental investigation of the flow downstream of a dysfunctional bileaflet mechanical aortic valve
Experimental investigation of the flow downstream of a dysfunctional bileaflet mechanical aortic valve
AbstractMechanical heart valve replacement is the preferred alternative in younger patients with severe symptomatic aortic valve disease. However, thrombus and pannus formations ar...
Statistical Characteristics of Flow Field through Open and Semi-Closed Bileaflet Mechanical Heart Valve
Statistical Characteristics of Flow Field through Open and Semi-Closed Bileaflet Mechanical Heart Valve
The formation of thrombi on the streamlined surface of the bileaflet mechanical heart valves is one of the main disadvantages of such valves. Thrombi block the valve leaflets and d...
Incidence, Outcomes and Predictors of Aortic Regurgitation After Transcatheter Aortic Valve Replacement in Al Azhar University Hospitals and National Heart Institute, Egypt
Incidence, Outcomes and Predictors of Aortic Regurgitation After Transcatheter Aortic Valve Replacement in Al Azhar University Hospitals and National Heart Institute, Egypt
Background: Aortic regurgitation (AR), which has ill-defined predictors and an unknown long-term influence on outcomes, is a significant transcatheter aortic valve replacement (TAV...
Abstract 4369914: Successful Transfemoral TAVR in a Patient with Chronic Aortic Dissection and Severe Aortic Insufficiency
Abstract 4369914: Successful Transfemoral TAVR in a Patient with Chronic Aortic Dissection and Severe Aortic Insufficiency
Transcatheter aortic valve replacement is a well-established treatment for high-surgical-risk patients with severe aortic disease, providing a less invasive alternative to traditio...
Increased aortic valve uptake of sodium fluoride is associated with higher cardiovascular risk: assessing pathophysiology with PET/CT
Increased aortic valve uptake of sodium fluoride is associated with higher cardiovascular risk: assessing pathophysiology with PET/CT
Abstract
Introduction
Positron emission tomography-computed tomography (PET-CT) with 18F-sodium fluoride (18F-NaF) has been used...
AORTIC ROOT ANEURYSM OR ECTASIA TREATED WITH AORTIC ROOT WRAPPED PROCEDURE
AORTIC ROOT ANEURYSM OR ECTASIA TREATED WITH AORTIC ROOT WRAPPED PROCEDURE
Objectives
To develop a relatively simple and effective and less risk operation, aortic root wrapped procedure, to treat with aortic root aneurysm or ectasia.
...
Aortic root wrapped procedure
Aortic root wrapped procedure
Objective
To develop a relatively simple and effective and less risk operation, aortic root wrapped procedure, to treat with aortic root aneurysm or ectasia.
...
Blood pressure, hypertension, and the risk of aortic aneurysm in the UK Biobank
Blood pressure, hypertension, and the risk of aortic aneurysm in the UK Biobank
Abstract
Background
Although an association between elevated blood pressure and risk of aortic aneurysm is established, f...