Investigating focal spot position drift in a mobile imaging system equipped with a monobloc‐based x‐ray generator

View through CrossRef
AbstractBackgroundMisalignment or double‐contouring artifacts can appear in high‐resolution 3D cone beam computed tomography (CBCT) images, potentially indicating geometric accuracy issues in the projection data. Such artifacts may go unnoticed in low‐resolution images and could be associated with changes in the focal spot (FS) position.PurposeHigh‐resolution 3D‐CBCT imaging by a mobile imaging device with a large gantry clearance offers more versatility for clinical workflows in image‐guided brachytherapy (IGBT), intraoperative radiation therapy (IORT), and spinal, as well as maxillofacial surgery. However, misalignment or double‐contouring artifacts hinder workflow advancements in these domains. This paper introduces intrinsic calibration and geometrical correction methods as extensions to a well‐established technique for addressing geometrical deviations resulting from factors such as gravity or mechanical inconsistencies. These extensions cover shifts and drifts of the FS depending on FS size selection, temperature, tube current, and tube potential. The proposed methods effectively mitigate artifacts in high‐resolution CBCT images stemming from geometrical inaccuracies in projection data, without requiring additional equipment like a pinhole device.MethodsGeometrical offsets and drifts of the x‐ray tube FS were characterized on a mobile multi‐purpose imaging system, the ImagingRing‐m. A pinhole‐like experiment was simulated by adjusting the movable collimation unit to a small rectangular aperture within the FS size range. The influence of filament selection, that is, FS size, temperature, the relatively low tube currents, as well as tube potential settings have been studied on two different monobloc types sharing the same x‐ray tube insert. The Catphan 504 and an Alderson head phantom were used to assess resulting image artifacts.ResultsSwitching the FS size to one different from what was used for geometrical (gravitation, mechanical variations) calibration induced the most notable position changes of the x‐ray FS, resulting in double‐contouring artifacts and blurring of high‐resolution 3D‐CBCT images. Incorporating these shifts into a geometrical correction method effectively minimized these artifacts. Thermal drifts exhibited the second largest geometrical changes, comparable to FS size shifts across the thermal operating conditions of the x‐ray system. The proposed thermal drift compensation markedly reduced thermal drift effects. Tube current and potential had little impact within the range of available tube currents, eliminating the need for compensation in current applications.ConclusionsAugmenting the geometrical calibration pipeline with proposed FS drift compensations yielded significant enhancements in image quality for high‐resolution reconstructions. While compensation for thermal effects posed challenges, it proved achievable. The roles of tube current and potential were found to be negligible.
Title: Investigating focal spot position drift in a mobile imaging system equipped with a monobloc‐based x‐ray generator
Description:
AbstractBackgroundMisalignment or double‐contouring artifacts can appear in high‐resolution 3D cone beam computed tomography (CBCT) images, potentially indicating geometric accuracy issues in the projection data.
Such artifacts may go unnoticed in low‐resolution images and could be associated with changes in the focal spot (FS) position.
PurposeHigh‐resolution 3D‐CBCT imaging by a mobile imaging device with a large gantry clearance offers more versatility for clinical workflows in image‐guided brachytherapy (IGBT), intraoperative radiation therapy (IORT), and spinal, as well as maxillofacial surgery.
However, misalignment or double‐contouring artifacts hinder workflow advancements in these domains.
This paper introduces intrinsic calibration and geometrical correction methods as extensions to a well‐established technique for addressing geometrical deviations resulting from factors such as gravity or mechanical inconsistencies.
These extensions cover shifts and drifts of the FS depending on FS size selection, temperature, tube current, and tube potential.
The proposed methods effectively mitigate artifacts in high‐resolution CBCT images stemming from geometrical inaccuracies in projection data, without requiring additional equipment like a pinhole device.
MethodsGeometrical offsets and drifts of the x‐ray tube FS were characterized on a mobile multi‐purpose imaging system, the ImagingRing‐m.
A pinhole‐like experiment was simulated by adjusting the movable collimation unit to a small rectangular aperture within the FS size range.
The influence of filament selection, that is, FS size, temperature, the relatively low tube currents, as well as tube potential settings have been studied on two different monobloc types sharing the same x‐ray tube insert.
The Catphan 504 and an Alderson head phantom were used to assess resulting image artifacts.
ResultsSwitching the FS size to one different from what was used for geometrical (gravitation, mechanical variations) calibration induced the most notable position changes of the x‐ray FS, resulting in double‐contouring artifacts and blurring of high‐resolution 3D‐CBCT images.
Incorporating these shifts into a geometrical correction method effectively minimized these artifacts.
Thermal drifts exhibited the second largest geometrical changes, comparable to FS size shifts across the thermal operating conditions of the x‐ray system.
The proposed thermal drift compensation markedly reduced thermal drift effects.
Tube current and potential had little impact within the range of available tube currents, eliminating the need for compensation in current applications.
ConclusionsAugmenting the geometrical calibration pipeline with proposed FS drift compensations yielded significant enhancements in image quality for high‐resolution reconstructions.
While compensation for thermal effects posed challenges, it proved achievable.
The roles of tube current and potential were found to be negligible.

Related Results

Mobile phone usage for m‐learning: comparing heavy and light mobile phone users
Mobile phone usage for m‐learning: comparing heavy and light mobile phone users
PurposeMobile technologies offer the opportunity to embed learning in a natural environment. The objective of the study is to examine how the usage of mobile phones for m‐learning ...
Analisis Penyebab Pecahnya Cylinder Liner pada Generator Engine di Kapal MV. Kali Mas
Analisis Penyebab Pecahnya Cylinder Liner pada Generator Engine di Kapal MV. Kali Mas
Globalization has spurred the growth of the transportation sector. In this context, maintenance of generators on ship machinery is essential to ensure smooth ship operations. The g...
ACM SIGCOMM computer communication review
ACM SIGCOMM computer communication review
At some point in the future, how far out we do not exactly know, wireless access to the Internet will outstrip all other forms of access bringing the freedom of mobility to the way...
Low-Pass Filters for a Temperature Drift Correction Method for Electromagnetic Induction Systems
Low-Pass Filters for a Temperature Drift Correction Method for Electromagnetic Induction Systems
Electromagnetic induction (EMI) systems are used for mapping the soil’s electrical conductivity in near-surface applications. EMI measurements are commonly affected by time-varying...
A new sea ice state dependent parameterization for the free drift of sea ice
A new sea ice state dependent parameterization for the free drift of sea ice
Abstract. Free drift estimates of sea ice motion are necessary to produce a seamless observational record combining buoy and satellite-derived sea ice motion vectors. We develop a ...
Analisa Pengaruh Tegangan Harmonik Terhadap Regulasi Tegangan Eksitasi Generator Satu Fasa
Analisa Pengaruh Tegangan Harmonik Terhadap Regulasi Tegangan Eksitasi Generator Satu Fasa
Esensinya setiap generator listrik satu fasa maupun tiga fasa telah dilengkapi dengan sistem eksitasi. Sistem eksitasi generator ada tiga, yaitu sistem eksitasi statis, dinamis, da...
Pendayagunaan Energi Matahari sebagai Sumber Energi Listrik Alternatif Menggunakan Generator Termoelektrik
Pendayagunaan Energi Matahari sebagai Sumber Energi Listrik Alternatif Menggunakan Generator Termoelektrik
Abstrak Energi surya adalah energi terbarukan yang potensinya besar untuk dimanfaatkan. Bentuk energi surya yang telah lama dimanfaatkan manusia adalah energi termalnya. Mulai dar...
The Stokes drift in ocean surface drift prediction
The Stokes drift in ocean surface drift prediction
<p>Ocean surface drift forecasts are essential for numerous applications. It is a central asset in search and rescue and oil spill response operations, but it is also...

Back to Top