Investigating focal spot position drift in a mobile imaging system equipped with a monobloc‐based x‐ray generator
View through CrossRef
AbstractBackgroundMisalignment or double‐contouring artifacts can appear in high‐resolution 3D cone beam computed tomography (CBCT) images, potentially indicating geometric accuracy issues in the projection data. Such artifacts may go unnoticed in low‐resolution images and could be associated with changes in the focal spot (FS) position.PurposeHigh‐resolution 3D‐CBCT imaging by a mobile imaging device with a large gantry clearance offers more versatility for clinical workflows in image‐guided brachytherapy (IGBT), intraoperative radiation therapy (IORT), and spinal, as well as maxillofacial surgery. However, misalignment or double‐contouring artifacts hinder workflow advancements in these domains. This paper introduces intrinsic calibration and geometrical correction methods as extensions to a well‐established technique for addressing geometrical deviations resulting from factors such as gravity or mechanical inconsistencies. These extensions cover shifts and drifts of the FS depending on FS size selection, temperature, tube current, and tube potential. The proposed methods effectively mitigate artifacts in high‐resolution CBCT images stemming from geometrical inaccuracies in projection data, without requiring additional equipment like a pinhole device.MethodsGeometrical offsets and drifts of the x‐ray tube FS were characterized on a mobile multi‐purpose imaging system, the ImagingRing‐m. A pinhole‐like experiment was simulated by adjusting the movable collimation unit to a small rectangular aperture within the FS size range. The influence of filament selection, that is, FS size, temperature, the relatively low tube currents, as well as tube potential settings have been studied on two different monobloc types sharing the same x‐ray tube insert. The Catphan 504 and an Alderson head phantom were used to assess resulting image artifacts.ResultsSwitching the FS size to one different from what was used for geometrical (gravitation, mechanical variations) calibration induced the most notable position changes of the x‐ray FS, resulting in double‐contouring artifacts and blurring of high‐resolution 3D‐CBCT images. Incorporating these shifts into a geometrical correction method effectively minimized these artifacts. Thermal drifts exhibited the second largest geometrical changes, comparable to FS size shifts across the thermal operating conditions of the x‐ray system. The proposed thermal drift compensation markedly reduced thermal drift effects. Tube current and potential had little impact within the range of available tube currents, eliminating the need for compensation in current applications.ConclusionsAugmenting the geometrical calibration pipeline with proposed FS drift compensations yielded significant enhancements in image quality for high‐resolution reconstructions. While compensation for thermal effects posed challenges, it proved achievable. The roles of tube current and potential were found to be negligible.
Title: Investigating focal spot position drift in a mobile imaging system equipped with a monobloc‐based x‐ray generator
Description:
AbstractBackgroundMisalignment or double‐contouring artifacts can appear in high‐resolution 3D cone beam computed tomography (CBCT) images, potentially indicating geometric accuracy issues in the projection data.
Such artifacts may go unnoticed in low‐resolution images and could be associated with changes in the focal spot (FS) position.
PurposeHigh‐resolution 3D‐CBCT imaging by a mobile imaging device with a large gantry clearance offers more versatility for clinical workflows in image‐guided brachytherapy (IGBT), intraoperative radiation therapy (IORT), and spinal, as well as maxillofacial surgery.
However, misalignment or double‐contouring artifacts hinder workflow advancements in these domains.
This paper introduces intrinsic calibration and geometrical correction methods as extensions to a well‐established technique for addressing geometrical deviations resulting from factors such as gravity or mechanical inconsistencies.
These extensions cover shifts and drifts of the FS depending on FS size selection, temperature, tube current, and tube potential.
The proposed methods effectively mitigate artifacts in high‐resolution CBCT images stemming from geometrical inaccuracies in projection data, without requiring additional equipment like a pinhole device.
MethodsGeometrical offsets and drifts of the x‐ray tube FS were characterized on a mobile multi‐purpose imaging system, the ImagingRing‐m.
A pinhole‐like experiment was simulated by adjusting the movable collimation unit to a small rectangular aperture within the FS size range.
The influence of filament selection, that is, FS size, temperature, the relatively low tube currents, as well as tube potential settings have been studied on two different monobloc types sharing the same x‐ray tube insert.
The Catphan 504 and an Alderson head phantom were used to assess resulting image artifacts.
ResultsSwitching the FS size to one different from what was used for geometrical (gravitation, mechanical variations) calibration induced the most notable position changes of the x‐ray FS, resulting in double‐contouring artifacts and blurring of high‐resolution 3D‐CBCT images.
Incorporating these shifts into a geometrical correction method effectively minimized these artifacts.
Thermal drifts exhibited the second largest geometrical changes, comparable to FS size shifts across the thermal operating conditions of the x‐ray system.
The proposed thermal drift compensation markedly reduced thermal drift effects.
Tube current and potential had little impact within the range of available tube currents, eliminating the need for compensation in current applications.
ConclusionsAugmenting the geometrical calibration pipeline with proposed FS drift compensations yielded significant enhancements in image quality for high‐resolution reconstructions.
While compensation for thermal effects posed challenges, it proved achievable.
The roles of tube current and potential were found to be negligible.
Related Results
Applying a user-centered approach to evaluate the usability of a mobile application for health professionals in home care services (Preprint)
Applying a user-centered approach to evaluate the usability of a mobile application for health professionals in home care services (Preprint)
BACKGROUND
Mobile health (mHealth), or the use of mobile devices in medicine and health, is a sub-category of e-health. mHealth interventions are designed t...
Analisis Penyebab Pecahnya Cylinder Liner pada Generator Engine di Kapal MV. Kali Mas
Analisis Penyebab Pecahnya Cylinder Liner pada Generator Engine di Kapal MV. Kali Mas
Globalization has spurred the growth of the transportation sector. In this context, maintenance of generators on ship machinery is essential to ensure smooth ship operations. The g...
A Comparative Study of Pyrenophora Teres Drechs. F. Teres and Pyrenophora Teres Drechs. F. Maculata Smedeg. Cause of "Net" and "Spot" Type Net Blotch Respectively on New Zealand Barley
A Comparative Study of Pyrenophora Teres Drechs. F. Teres and Pyrenophora Teres Drechs. F. Maculata Smedeg. Cause of "Net" and "Spot" Type Net Blotch Respectively on New Zealand Barley
<p>Net blotch is caused by Pyrenophora teres Drechs. (stat. conid. Drechslera teres (Sacc.) Shoem., syn. Helminthosporium teres Sacc). P. teres produces symptoms which appear...
Mobile phone usage for m‐learning: comparing heavy and light mobile phone users
Mobile phone usage for m‐learning: comparing heavy and light mobile phone users
PurposeMobile technologies offer the opportunity to embed learning in a natural environment. The objective of the study is to examine how the usage of mobile phones for m‐learning ...
ACM SIGCOMM computer communication review
ACM SIGCOMM computer communication review
At some point in the future, how far out we do not exactly know, wireless access to the Internet will outstrip all other forms of access bringing the freedom of mobility to the way...
Analysis of a Mobile Learning Adoption Model for Learning Improvement Based on Students’ Perception
Analysis of a Mobile Learning Adoption Model for Learning Improvement Based on Students’ Perception
Aim/Purpose: This identifies the factors that influence the application of mobile learning in order to improve the student learning process at universities in Indonesia based on th...
Low-Pass Filters for a Temperature Drift Correction Method for Electromagnetic Induction Systems
Low-Pass Filters for a Temperature Drift Correction Method for Electromagnetic Induction Systems
Electromagnetic induction (EMI) systems are used for mapping the soil’s electrical conductivity in near-surface applications. EMI measurements are commonly affected by time-varying...
A new sea ice state dependent parameterization for the free drift of sea ice
A new sea ice state dependent parameterization for the free drift of sea ice
Abstract. Free drift estimates of sea ice motion are necessary to produce a seamless observational record combining buoy and satellite-derived sea ice motion vectors. We develop a ...