A prospective randomised comparison of two different stenting sutures when using the Baerveldt 101–350 aqueous shunt – Lab transition to clinical process

View through CrossRef
Abstract Purpose The Baerveldt glaucoma drainage implant (BGI 101–350) is one of the most effective treatment options for managing refractory glaucoma. To mitigate early hypotony in non‐valved implants, intraluminal occluding sutures (IOSs) are commonly used. This study compares the swelling behavior of two widely used IOSs: 3–0 nylon (Supramid Extra) and 3–0 polypropylene (Prolene). Long‐term clinical outcomes associated with each suture material are also evaluated. Methods A prospective randomised interventional study was conducted involving 94 patients undergoing BGI 101–350 implantation, divided into two groups: Supramid ( n  = 48) and Prolene ( n  = 46). An experimental swelling model using confocal microscopy and ImageJ software quantified the hygroscopic behavior of each suture material by determining the intraluminal gaps area before and after wetting. Postoperative intraocular pressure (IOP), medication use and removal timing were analyzed over 60 months using independent‐samples t ‐tests, while categorical variables, including hypotony rates and the need for suture adjustment, were compared using χ 2 tests. Results The experimental model demonstrated ~60% swelling in Supramid vs. ~3% in Prolene IOS. Clinically, both groups achieved comparable IOP control at 6, 12 and 60 months (mean IOP at 60 months: Supramid 11 ± 1.75 mmHg, Prolene 11 ± 1.37 mmHg; p  > 0.05). The Prolene group required markedly fewer postoperative interventions: no suture adjustments compared with 10 in the Supramid group ( p  = 0.001), and suture removal occurred significantly later (mean 8.9 vs. 4.96 months; p  < 0.001). Transient hypotony following suture removal occurred in 6.25% of Supramid cases and in none with Prolene. Conclusions 3–0 Prolene offers better predictability and repeatability, earlier IOP control, delayed and safer suture removal and reduced need for early interventions while maintaining equivalent long‐term IOP control.
Title: A prospective randomised comparison of two different stenting sutures when using the Baerveldt 101–350 aqueous shunt – Lab transition to clinical process
Description:
Abstract Purpose The Baerveldt glaucoma drainage implant (BGI 101–350) is one of the most effective treatment options for managing refractory glaucoma.
To mitigate early hypotony in non‐valved implants, intraluminal occluding sutures (IOSs) are commonly used.
This study compares the swelling behavior of two widely used IOSs: 3–0 nylon (Supramid Extra) and 3–0 polypropylene (Prolene).
Long‐term clinical outcomes associated with each suture material are also evaluated.
Methods A prospective randomised interventional study was conducted involving 94 patients undergoing BGI 101–350 implantation, divided into two groups: Supramid ( n  = 48) and Prolene ( n  = 46).
An experimental swelling model using confocal microscopy and ImageJ software quantified the hygroscopic behavior of each suture material by determining the intraluminal gaps area before and after wetting.
Postoperative intraocular pressure (IOP), medication use and removal timing were analyzed over 60 months using independent‐samples t ‐tests, while categorical variables, including hypotony rates and the need for suture adjustment, were compared using χ 2 tests.
Results The experimental model demonstrated ~60% swelling in Supramid vs.
~3% in Prolene IOS.
Clinically, both groups achieved comparable IOP control at 6, 12 and 60 months (mean IOP at 60 months: Supramid 11 ± 1.
75 mmHg, Prolene 11 ± 1.
37 mmHg; p  > 0.
05).
The Prolene group required markedly fewer postoperative interventions: no suture adjustments compared with 10 in the Supramid group ( p  = 0.
001), and suture removal occurred significantly later (mean 8.
9 vs.
4.
96 months; p  < 0.
001).
Transient hypotony following suture removal occurred in 6.
25% of Supramid cases and in none with Prolene.
Conclusions 3–0 Prolene offers better predictability and repeatability, earlier IOP control, delayed and safer suture removal and reduced need for early interventions while maintaining equivalent long‐term IOP control.

Related Results

Pohjois-Ruotsin suomalaisten parissa
Pohjois-Ruotsin suomalaisten parissa
Kielenainekset(k)renkku (kieli: suomi, sivulla: 101)ahma (kieli: suomi, sivulla: 101)airokas (kieli: suomi, sivulla: 100)arina (kieli: suomi, sivulla: 101)arvaus (kieli: suomi, siv...
David Skogman Pohjois-Ruotsin suomalaisten parissa [David Skogmanin päiväkirja, muistiinpanoja kielestä, kielennäytteitä]
David Skogman Pohjois-Ruotsin suomalaisten parissa [David Skogmanin päiväkirja, muistiinpanoja kielestä, kielennäytteitä]
Kirja-arvioSkogman, David: Pohjois-Ruotsin suomalaisten parissa [David Skogmanin päiväkirja, muistiinpanoja kielestä, kielennäytteitä]Kielenainekset(k)renkku (kieli: suomi, sivulla...
Pooled Data Analysis of the Ahmed Baerveldt Comparison Study and the Ahmed Versus Baerveldt Study
Pooled Data Analysis of the Ahmed Baerveldt Comparison Study and the Ahmed Versus Baerveldt Study
Data from 2 independent randomized clinical trials, the Ahmed Baerveldt Comparison (ABC) Study and the Ahmed Versus Baerveldt (AVB) Study, were pooled for analysis to determine whe...
Current therapeutic strategies for erectile function recovery after radical prostatectomy – literature review and meta-analysis
Current therapeutic strategies for erectile function recovery after radical prostatectomy – literature review and meta-analysis
Radical prostatectomy is the most commonly performed treatment option for localised prostate cancer. In the last decades the surgical technique has been improved and modified in or...
XEN-augmented Baerveldt: A New Surgical Technique for Refractory Glaucoma
XEN-augmented Baerveldt: A New Surgical Technique for Refractory Glaucoma
Background: Glaucoma drainage devices have traditionally been reserved for patients with refractory glaucoma. However, these devices are prone to various sight-threaten...
Uncommon complications of ventriculoperitoneal shunt surgery: review of four cases and literature review
Uncommon complications of ventriculoperitoneal shunt surgery: review of four cases and literature review
Abstract Background Ventriculoperitoneal shunt is one of the most popular cerebrospinal fluid diversion procedures worldwide. Complications are common, but uncommon complications a...

Back to Top