The dissolution of urinary catheter encrustation

View through CrossRef
Objectives  To identify the optimum volume of acidic bladder washout solution to dissolve catheter encrustations and to compare the effectiveness of different bladder washout delivery devices. Materials and methods  Urinary catheter encrustation was generated in vitro using a model of the catheterized bladder adapted from previous work. An acidic bladder washout solution (Suby G) was applied via the catheter in the model and retained for 15 min. The amount of encrusting material dissolved was measured by colorimetric analysis of the magnesium and calcium content of the solution returned after the washout procedure. The relative effectiveness of different volumes of washout solution and different washout delivery devices (Optiflow, Bard Ltd, Crawley; UroTainer, B/Braun Medical Ltd, Aylesbury; and a bladder syringe) in dissolving catheter encrustations was compared to the ‘standardized’ conditions commonly used in practice. Results  There was no statistically significant difference between washouts with 100 mL and washouts with 50 mL for any of the delivery devices tested. Gentle agitation with 50 mL Suby G did not significantly improve the dissolution of encrustation but the total amount of material dissolved when two washouts with 50 mL Suby G were used sequentially was significantly better than a single washout with either 100 mL or 50 mL. Conclusions  Under controlled laboratory conditions, smaller volumes of acidic bladder washout solution (50 mL) are as effective as the 100 mL commonly used in practice, but two sequential washouts with 50 mL are more effective than a single washout. The newly designed Optiflow delivery device is at least as effective as the other devices tested. This study provides new evidence which may contribute to decision‐making by practitioners in relation to cath‐eter care.
Title: The dissolution of urinary catheter encrustation
Description:
Objectives  To identify the optimum volume of acidic bladder washout solution to dissolve catheter encrustations and to compare the effectiveness of different bladder washout delivery devices.
Materials and methods  Urinary catheter encrustation was generated in vitro using a model of the catheterized bladder adapted from previous work.
An acidic bladder washout solution (Suby G) was applied via the catheter in the model and retained for 15 min.
The amount of encrusting material dissolved was measured by colorimetric analysis of the magnesium and calcium content of the solution returned after the washout procedure.
The relative effectiveness of different volumes of washout solution and different washout delivery devices (Optiflow, Bard Ltd, Crawley; UroTainer, B/Braun Medical Ltd, Aylesbury; and a bladder syringe) in dissolving catheter encrustations was compared to the ‘standardized’ conditions commonly used in practice.
Results  There was no statistically significant difference between washouts with 100 mL and washouts with 50 mL for any of the delivery devices tested.
Gentle agitation with 50 mL Suby G did not significantly improve the dissolution of encrustation but the total amount of material dissolved when two washouts with 50 mL Suby G were used sequentially was significantly better than a single washout with either 100 mL or 50 mL.
Conclusions  Under controlled laboratory conditions, smaller volumes of acidic bladder washout solution (50 mL) are as effective as the 100 mL commonly used in practice, but two sequential washouts with 50 mL are more effective than a single washout.
The newly designed Optiflow delivery device is at least as effective as the other devices tested.
This study provides new evidence which may contribute to decision‐making by practitioners in relation to cath‐eter care.

Related Results

Study of factors influencing the encrustation of indwelling catheters: prospective case series
Study of factors influencing the encrustation of indwelling catheters: prospective case series
Abstract Background Long-term catheterisation is frequently necessary for patients with retention of urine or intractable incontinence who are unfit...
Recent Analytical Techniques for Encrustation in Ureteral Stents: A Review
Recent Analytical Techniques for Encrustation in Ureteral Stents: A Review
The purpose of this review is to explore the issue of ureteral stent encrustation, a common complication following stent placement, and to examine recent advancements in the litera...
Sodium Citrate Versus Heparin Locking Solutions for Tunneled Hemodialysis Catheters
Sodium Citrate Versus Heparin Locking Solutions for Tunneled Hemodialysis Catheters
Introduction: Hemodialysis catheter dysfunction compromises dialysis adequacy and may limit catheter survival. Strategies to prevent hemodialysis catheter dysfunction include cathe...
Does Immediate Removal of Urinary Catheter Prevent Urinary Morbidities Following Cesarean Section: A Prospective Randomized Study
Does Immediate Removal of Urinary Catheter Prevent Urinary Morbidities Following Cesarean Section: A Prospective Randomized Study
Objective: Urinary catheters are known cause of urinary morbidities. The longer the catheter is retained, the greater the risk for contamination and infection. An increasing body o...

Back to Top