Javascript must be enabled to continue!
Validity of the Fitbit activity tracker for measuring steps in community-dwelling older adults
View through CrossRef
Background
Commercially available activity monitors, such as the Fitbit, may encourage physical activity. However, the accuracy of the Fitbit in older adults remains unknown. This study aimed to determine (1) the criterion validity of Fitbit step counts compared to visual count and ActiGraph accelerometer step counts and (2) the accuracy of ActiGraph step counts compared to visual count in community-dwelling older people.
Methods
Thirty-two community-dwelling adults aged over 60 wore Fitbit and ActiGraph devices simultaneously during a 2 min walk test (2MWT) and then during waking hours over a 7-day period. A physiotherapist counted the steps taken during the 2MWT.
Results
There was excellent agreement between Fitbit and visually counted steps (intraclass correlation coefficient (ICC2,1)=0.88, 95% CI 0.76 to 0.94) from the 2MWT, and good agreement between Fitbit and ActiGraph (ICC2,1=0.66, 95% CI 0.41 to 0.82), and between ActiGraph and visually counted steps (ICC2,1=0.60, 95% CI 0.33 to 0.79). There was excellent agreement between the Fitbit and ActiGraph in average steps/day over 7 days (ICC2,1=0.94, 95% CI 0.88 to 0.97). Percentage agreement was closest for Fitbit steps compared to visual count (mean 0%, SD 4%) and least for Fitbit average steps/day compared to the ActiGraph (mean 13%, SD 25%).
Conclusions
The Fitbit accurately tracked steps during the 2MWT, but the ActiGraph appeared to underestimate steps. There was strong agreement between Fitbit and ActiGraph counted steps. The Fitbit tracker is sufficiently accurate to be used among community-dwelling older adults to monitor and give feedback on step counts.
Title: Validity of the Fitbit activity tracker for measuring steps in community-dwelling older adults
Description:
Background
Commercially available activity monitors, such as the Fitbit, may encourage physical activity.
However, the accuracy of the Fitbit in older adults remains unknown.
This study aimed to determine (1) the criterion validity of Fitbit step counts compared to visual count and ActiGraph accelerometer step counts and (2) the accuracy of ActiGraph step counts compared to visual count in community-dwelling older people.
Methods
Thirty-two community-dwelling adults aged over 60 wore Fitbit and ActiGraph devices simultaneously during a 2 min walk test (2MWT) and then during waking hours over a 7-day period.
A physiotherapist counted the steps taken during the 2MWT.
Results
There was excellent agreement between Fitbit and visually counted steps (intraclass correlation coefficient (ICC2,1)=0.
88, 95% CI 0.
76 to 0.
94) from the 2MWT, and good agreement between Fitbit and ActiGraph (ICC2,1=0.
66, 95% CI 0.
41 to 0.
82), and between ActiGraph and visually counted steps (ICC2,1=0.
60, 95% CI 0.
33 to 0.
79).
There was excellent agreement between the Fitbit and ActiGraph in average steps/day over 7 days (ICC2,1=0.
94, 95% CI 0.
88 to 0.
97).
Percentage agreement was closest for Fitbit steps compared to visual count (mean 0%, SD 4%) and least for Fitbit average steps/day compared to the ActiGraph (mean 13%, SD 25%).
Conclusions
The Fitbit accurately tracked steps during the 2MWT, but the ActiGraph appeared to underestimate steps.
There was strong agreement between Fitbit and ActiGraph counted steps.
The Fitbit tracker is sufficiently accurate to be used among community-dwelling older adults to monitor and give feedback on step counts.
Related Results
Evaluating the Science to Inform the Physical Activity Guidelines for Americans Midcourse Report
Evaluating the Science to Inform the Physical Activity Guidelines for Americans Midcourse Report
Abstract
The Physical Activity Guidelines for Americans (Guidelines) advises older adults to be as active as possible. Yet, despite the well documented benefits of physical a...
Is a Fitbit a Diary? Self-Tracking and Autobiography
Is a Fitbit a Diary? Self-Tracking and Autobiography
Data becomes something of a mirror in which people see themselves reflected. (Sorapure 270)In a 2014 essay for The New Yorker, the humourist David Sedaris recounts an obsession spu...
Physical Activity in Older Adults: Benefits, Risks of Inactivity, and Motivational Factors
Physical Activity in Older Adults: Benefits, Risks of Inactivity, and Motivational Factors
Background: Physical activity is essential for promoting health, autonomy and quality of life in older adults. However, adherence to exercise remains low worldwide. Motivation is r...
Exploring the effects of fitbit incentive on treatment outcomes in veterans undergoing intensive pain rehabilitation program
Exploring the effects of fitbit incentive on treatment outcomes in veterans undergoing intensive pain rehabilitation program
Abstract
Objective
This study compares clinical pain outcomes between patients in a pain treatment program that received a Fitbit, to patients that ...
Validation of the Fitbit Wireless Activity Tracker for Prediction of Energy Expenditure
Validation of the Fitbit Wireless Activity Tracker for Prediction of Energy Expenditure
Objective:The purpose of this study was to examine the accuracy of the Fitbit wireless activity tracker in assessing energy expenditure (EE) for different activities.Methods:Twenty...
Depression in geriatrics: a systematic review and meta-analysis of prevalence and risk factors in Egypt
Depression in geriatrics: a systematic review and meta-analysis of prevalence and risk factors in Egypt
Abstract
Background
Depression is the most common psychiatric disorder in older adults, even though it is commonly misdiagnosed and undertreated, le...
Experience of fall
Experience of fall
Background: Community-dwelling older adults' falls represent a significant public health problem, and prevention is an important step to reducing healthcare costs and improving old...
Sensor-Based Frailty Assessment Using Fitbit
Sensor-Based Frailty Assessment Using Fitbit
This study evaluates the reliability of Fitbit in assessing frailty based on motor and heart rate (HR) parameters through a validated upper extremity function (UEF) test, which inv...